![Tektronix P7313 User Manual Download Page 39](http://html.mh-extra.com/html/tektronix/p7313/p7313_user-manual_1079066039.webp)
Operating Basics
Z-Active Differential Probe Family User Manual
27
Wire Replacement (P7313 Only)
If you need to resolder a HBW Tip-Clip wire lead, refer to the
instructions in the Wire Replacement Kit. This kit is located inside
the Accessory Kit box that came with the probe. The Wire
Replacement Kit includes the high temperature solder recommended
for use in replacing wire leads on the HBW Tip-Clips. The kit also
includes both replacement 8 mil wire and 4 mil wire, which is used
with smaller circuit board traces and vias.
Typical Characteristic
Typical characteristics (Tables 1 and 2) describe typical but not
guaranteed performance
.
Table 1: Typical electrical characteristics
Characteristic
Description
Differential Signal Range
(DC Coupled)
P7380A, P7360A, & P7340A
±
1 V (5X)
±
2.5 V (25X)
P7313
±
0.625 V (5X)
±
2.0 V (25X)
Operating Voltage Window
P7380A, P7360A, & P7340A
--3.0 V to +5.0 V (5X and 25X)
P7313
--3.0 V to +4.0 V (5X and 25X)
Differential offset range
--3.0 V to +4.0 V
Maximum nondestructive
input voltage
1
±
15 V
(DC + peak AC)
between each input or
between either probe inputs and ground.
1
See the caution that follows.