Operating Basics
8
Z-Active Differential Probe Family User Manual
6.
Connect the BNC-to-Minigrabbers to the Long Flex, Small
Resistor, Tip-Clip assembly on the probe.
H
Connect the probe + input to the BNC red (+) terminal.
H
Connect the probe -- input to the adapter -- or ground.
Quick Tip:
The PPM203B Articulating Arm (used with the Probe
Arm Adapter) and PPM100 Positioner Arm are not required, but
may be used to support the probe. This relieves strain on the
Tip-Clip assembly and provides a stable waveform measurement.
See the
Accessories
section of this manual on page 40.
7.
Adjust the oscilloscope to display a stable waveform (we
recommend that you use the Autoset function). When you see a
stable square wave, check the amplitude.
The probe compensation signal amplitude and common mode
voltage is dependent on oscilloscope model. Check that the signal
amplitude on the oscilloscope and the common mode voltage
approximate those in the table:
P7313, P7380A, P7360A,
or P7340A probe set at
25X Attenuation
TDS6604
TDS7704
Signal amplitude
400 mV p-p
1.0 V p-p
V
CM
1.8 V
--500 mV