![Tektronix P7313 User Manual Download Page 28](http://html.mh-extra.com/html/tektronix/p7313/p7313_user-manual_1079066028.webp)
Operating Basics
16
Z-Active Differential Probe Family User Manual
Connect the probe to any oscilloscope that features the TekConnect
interface.
2.
Attach the probe to the Tip-Clip assembly.
CAUTION.
To prevent damage to the circuit board or circuit board
connections due to accidental movement of the probe and soldered
leads, it is recommended that the Tip-Clip assembly and probe be
secured to the circuit board using the provided Velcro fastening strap
and dots, and the Tip-Clip Tape
.
3.
Attach the Velcro fastening strap and dots to the probe and circuit
board.
CAUTION.
To avoid damage to the Tip-Clip or the circuit under test,
avoid applying excessive heat from the soldering iron. Use a low
wattage, temperature-controlled soldering iron and appropriately
sized soldering iron tip.
4.
Solder the Tip-Clip leads to test points on your circuit board. See
Quick Tips
.
Quick Tips
H
For optimum performance and signal integrity, keep the lead
length between the DUT (Device Under Test) and the Tip-Clip
resistor short, less than 0.050-in if possible, and the lead lengths
the same length.
H
Since the Tip-Clip loading is relatively light at about 0.1 pF, you
may be able to leave several Tip-Clip assemblies attached to your
circuit while you take measurements, depending on the
sensitivity of your circuit to loading. Tip-Clip loading of the
circuit under test is minimum when the probe is attached to the
Tip-Clip.