P6434 Probe Instructions
6.
Verify that the logic analyzer passes the power-on diagnostics.
7.
Go to the Setup menu for the logic analyzer module to which the probes are
connected.
8.
Click the Set Thresholds button to display the threshold settings. Set the
threshold for all channels to be 700 mV. Close the Probes Threshold dialog
box and click Set All.
9.
Click the Show Activity button to display the Activity Indicator dialog box.
10.
Visually verify that the Activity Indicator dialog box shows activity on all
probe channels that are connected to the test fixture (see Figure 34).
Note the signal activity for the clock, qualifier and data channels for the C3,
C2, C1, and C0 probe sections. Also note that there is no activity on other
probe sections because the probes are not connected to a signal source. (The
channels are all high.)
Signal activity
on all connected
channels
Stuck low
(No activity)
Stuck high
(No activity)
No activity
(high)
Figure 34: Example of the Activity Indicator dialog box
11.
Visually verify in the Activity Indicator dialog box that none of the
connected channels are stuck high or stuck low (see Figure 34).
CAUTION.
To prevent damage to the test fixture connector, grasp the mictor
connector on the test fixture when wiggling the probe tip end.