![Tektronix P6434 Instructions Manual Download Page 54](http://html1.mh-extra.com/html/tektronix/p6434/p6434_instructions-manual_1078102054.webp)
Reference
P6434 Probe Instructions
41
Functional Verification
This procedure checks the basic functionality of the probe by verifying that the
probes recognize signal activity at the probe tips. No calibration is necessary.
The functional verification procedure for the P6434 probe requires the following
test equipment:
H
Adjustment and verification test fixture (refer to the
Replaceable Parts List
section under
Optional Accessories
on page 48 for part number information)
H
Power supply for the test fixture (refer to the
Replaceable Parts List
section
under
Optional Accessories
on page 48 for part number information)
H
Logic analyzer mainframe
H
Logic analyzer module
To perform the functional verification procedure, follow these steps:
1.
Connect the module ends of the P6434 probe to the logic analyzer module.
2.
Connect the probe tip to J5 on the the test fixture as shown in Figure 33.
J5
J15
Figure 33: Jumper and connector locations on the test fixture
3.
Place J15 in the INT position to select the internal 50 MHz clock.
4.
Connect the power adapter to the test fixture and power on the test fixture.
5.
Power on the logic analyzer.