4
P6112 Instructions
Specifications
The characteristics listed in Table 1 apply to a P6112 Passive Probe
installed on a Tektronix TDS 220 oscilloscope unless otherwise
noted.
These specifications apply when the instrument has had a warm-up
period of at least 10 minutes and is in an environment that does not
exceed the limits described in Table 2.
Table 1: Electrical characteristics
Attenuation
(DC coupled system)
10X
"
2% at DC
Input resistance
(DC coupled system)
10 M
"
1.5%
Input capacitance
11.8 pF to 14.8 pF
(13.3 pF typical)
Compensation range
15 pF to 35 pF
Rise time
(DC coupled system)
3.5 ns (TDS 220)
5.8 ns (TDS 210)
Bandwidth
(–3 dB, DC coupled system)
DC to 100 MHz (TDS 220)
DC to 60 MHz (TDS 210)
Maximum Working (Nondestructive)
Input Voltage
(For frequency derating curve, refer
to Figure 1 on page 6)
Overvoltage Category
Category I
Category II
Maximum Voltage
300 V
RMS
300 V
RMS
Safety
UL3111–1, First Edition
CSA C22.2 No. 1010.1–92
EN61010–1
EN61010–2–031
Overvoltage Category II, Pollution Degree 2