![Tektronix LumaColor II J18 Instruction Manual Download Page 49](http://html1.mh-extra.com/html/tektronix/lumacolor-ii-j18/lumacolor-ii-j18_instruction-manual_1078036049.webp)
J18 Photometer Instruction Manual
3–7
Commands
The J18 LumaColor
II Photometer recognizes ten commands that
duplicate the function of the front panel controls. These commands
are summarized in Table 3–4.
NOTE
. The J18 is a case-sensitive device, so all RS-232 commands
should be made in capital letters (upper-case), for example: !NEW.
Table 3–4: J18 Command Set
Command
Function
!INI{CR}
Initializes the instrument. Puts the J18 in the initial
powerup state with no valid data for the attached sensor
head.
!ALT{CR}
Same as the ALT DISP button.
!BKL{CR}
Same as the BACKLIGHT button.
!DIF{CR}
Same as the DIFF/REF button.
!HLD{CR}
Same as the HOLD button.
!RGE{CR}
Same as the RANGE HI/LO button.
!STR{CR}
Same as the STORE button.
!UNT{CR}
Same as the UNITS button.
!NEW{CR}
Causes the J18 to report the last measurement taken.
!NEW n {CR}
Causes the J18 to report the specified number
(n = sample size) of measurements beginning with the
last one taken. The sample size may be from 1 to 255. A
value of n
≥
128 or greater causes the J18 to report
measurements continuously until another command is
sent.
J18 Photometer Instruction Manual
3–7
Commands
The J18 LumaColor
II Photometer recognizes ten commands that
duplicate the function of the front panel controls. These commands
are summarized in Table 3–4.
NOTE
. The J18 is a case-sensitive device, so all RS-232 commands
should be made in capital letters (upper-case), for example: !NEW.
Table 3–4: J18 Command Set
Command
Function
!INI{CR}
Initializes the instrument. Puts the J18 in the initial
powerup state with no valid data for the attached sensor
head.
!ALT{CR}
Same as the ALT DISP button.
!BKL{CR}
Same as the BACKLIGHT button.
!DIF{CR}
Same as the DIFF/REF button.
!HLD{CR}
Same as the HOLD button.
!RGE{CR}
Same as the RANGE HI/LO button.
!STR{CR}
Same as the STORE button.
!UNT{CR}
Same as the UNITS button.
!NEW{CR}
Causes the J18 to report the last measurement taken.
!NEW n {CR}
Causes the J18 to report the specified number
(n = sample size) of measurements beginning with the
last one taken. The sample size may be from 1 to 255. A
value of n
≥
128 or greater causes the J18 to report
measurements continuously until another command is
sent.
Summary of Contents for LumaColor II J18
Page 4: ......
Page 6: ......
Page 16: ...Preface x J18 Photometer Instruction Manual Preface x J18 Photometer Instruction Manual...
Page 17: ...Getting Started Getting Started...
Page 18: ......
Page 29: ...Operating Basics Operating Basics...
Page 30: ......
Page 41: ...Reference Reference...
Page 42: ......
Page 48: ...Syntax 3 6 J18 Photometer Instruction Manual Syntax 3 6 J18 Photometer Instruction Manual...
Page 50: ...Commands 3 8 J18 Photometer Instruction Manual Commands 3 8 J18 Photometer Instruction Manual...
Page 53: ...Specifications Specifications...
Page 54: ......
Page 60: ......
Page 61: ...Performance Verification Performance Verification...
Page 62: ......
Page 65: ...Maintenance Maintenance...
Page 66: ......
Page 73: ...Replaceable Parts Replaceable Parts...
Page 74: ......
Page 85: ...Appendices Appendices...
Page 86: ......
Page 103: ...Glossary and Index Glossary and Index...
Page 104: ......
Page 113: ......
Page 114: ......