Figure 3: LE display with Channel 1 enabled
3.
Set up the sampling oscilloscope as follows:
a.
Connect the channel 1 oscilloscope out put to the Ch 1 + output of the
linear equalizer.
b.
Connect the channel 2 oscilloscope out put to the Ch 1 - output of the
linear equalizer.
c.
Set the vertical controls of the oscilloscope to 100 mV per division.
d.
Set the horizontal controls of the oscilloscope to 10 ps per division.
e.
Center the traces horizontally on the display as shown in the following
figure.
Figure 4: Sampling oscilloscope display
Performance verification
LE160 & LE320 Specifications & Performance Verification
11