071368600 / September 2019 

11 

 

The 

CAUTION

 heading in the user documentation explains hazards that could damage the instrument. Such damage may 

invalidate the warranty. 

The 

CAUTION

 heading with the 

  symbol in the user documentation explains hazards that could result in moderate or minor 

injury or damage the instrument. Always read the associated information very carefully before performing the indicated 
procedure. Damage to the instrument may invalidate the warranty.   

Instrumentation and accessories shall not be connected to humans. 

Before performing any maintenance, disconnect the line cord and all test cables. 

To maintain protection from electric shock and fire, replacement components in mains circuits — including the power 
transformer, test leads, and input jacks — must be purchased from Keithley. Standard fuses with applicable national safety 
approvals may be used if the rating and type are the same. The detachable mains power cord provided with the instrument may 
only be replaced with a similarly rated power cord. Other components that are not safety-related may be purchased from other 
suppliers as long as they are equivalent to the original component (note that selected parts should be purchased only through 
Keithley to maintain accuracy and functionality of the product). If you are unsure about the applicability of a replacement 
component, call a Keithley office for information. 

Unless otherwise noted in product-specific literature, Keithley instruments are designed to operate indoors only, in the following 
environment: Altitude at or below 2,000 m (6,562 ft); temperature 0 °C to 50 °C (32 °F to 122 °F); and pollution degree 1 or 2. 

To clean an instrument, use a cloth dampened with deionized water or mild, water-based cleaner. Clean the exterior of the 
instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that 
consist of a circuit board with no case or chassis (e.g., a data acquisition board for installation into a computer) should never 
require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board 
should be returned to the factory for proper cleaning/servicing. 

Safety precaution revision as of June 2017. 

 

Summary of Contents for Keithley TRX-1100V-CONN

Page 1: ...SourceMeter Instrument It can be used for instrument testing that requires high voltage and high throughput combined with DC measurement capability and broad testing flexibility Figure 1 TRX 1100V CO...

Page 2: ...0 2 30 safety standards The purple marking also indicates that they can be attached to other connectors with purple markings such as the FORCE and SENSE connectors on the Model 2470 rear panel shown i...

Page 3: ...rmediate contact to outer body 2500 VRMS Operating environment 0 C to 50 C up to 70 relative humidity at 35 C SC 22 triaxial cable construction Figure 5 SC 22 triaxial cable dimensions SC 22 electrica...

Page 4: ...mensions to ensure the correct insertion depth in the connector Excessive exposed conductor length presents electrical shock hazards to equipment and personnel If the strip length is too short imprope...

Page 5: ...To prepare the SC 22 cable for assembly to the TRX 1100V CONN 1 Use the cable ringing tool or wire stripper to strip the cable to the proper dimensions Figure 7 Strip SC 22 cable 2 Remove the graphite...

Page 6: ...gure 9 Add hex nut and V gasket to triax cable 2 Install the V clamp as shown in the following figure Carefully seat the V clamp into the V gasket when inserting The sharp edge of the V clamp must spl...

Page 7: ...lamp onto the V clamp Figure 12 Add front clamp 5 Install the gasket clamp insulator and inner clamp as shown in the following figure Figure 13 Add gasket clamp insulator and inner clamp 6 Flare the e...

Page 8: ...ntact Crimp the center contact to the center conductor cable as shown in the following figure Figure 15 Add slotted clamp support insulator and center contact 8 Slide the outer contact over the center...

Page 9: ...ns 071368600 September 2019 9 9 Insert the terminated cable assembly into the rear of the connector body as shown in the following figure 10 Tighten the cable nut with a wrench to 3 39 Newton meters 3...

Page 10: ...ts will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to mains Exercise extreme caution when a...

Page 11: ...r components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component note that selected parts should be purchased only through Keithl...

Reviews: