Model TRX-1100V-CONN 3-Slot Male HV Triaxial Cable Connector Instructions 

071368600 / September 2019 

Triaxial cable connector electrical characteristics 

Characteristic impedance: 

Nonconstant

Insulation resistance:

4 x 10 Ω

15

 minimum

Voltage rating:

Center contact to intermediate 500 V

RMS

Intermediate contact to outer body 1000 V

RMS

Dielectric withstanding voltage:

Center contact to intermediate 1500 V

RMS

Intermediate contact to outer body 2500 V

RMS

Operating environment: 0 °C to 50 

°C up to 70% relative humidity at ≤ 35

 °C

SC-22 triaxial cable construction 

Figure 5: SC-22 triax

ial

 cable dimensions 

SC-22 electrical characteristics 

Characteristic impedance:

40 Ω nominal center conductor to inner shield

15.3 Ω nominal shield to shield

Center conductor resistance:

< 0.1 Ω per 0.3048 cm (1 foot)

Voltage breakdown:

2655 volt DC minimum between shields

Flexural noise:

Peak-to-peak 

 7 mV

Leakage resistance (3.048 m [3 ft] cable length):

Center conductor to inner shield > 1.2 x 10

11 

Ω at 1200

 

V

Inner shield to outer shield > 1.6 x 10

11 

Ω at 1600 V

Summary of Contents for Keithley TRX-1100V-CONN

Page 1: ...SourceMeter Instrument It can be used for instrument testing that requires high voltage and high throughput combined with DC measurement capability and broad testing flexibility Figure 1 TRX 1100V CO...

Page 2: ...0 2 30 safety standards The purple marking also indicates that they can be attached to other connectors with purple markings such as the FORCE and SENSE connectors on the Model 2470 rear panel shown i...

Page 3: ...rmediate contact to outer body 2500 VRMS Operating environment 0 C to 50 C up to 70 relative humidity at 35 C SC 22 triaxial cable construction Figure 5 SC 22 triaxial cable dimensions SC 22 electrica...

Page 4: ...mensions to ensure the correct insertion depth in the connector Excessive exposed conductor length presents electrical shock hazards to equipment and personnel If the strip length is too short imprope...

Page 5: ...To prepare the SC 22 cable for assembly to the TRX 1100V CONN 1 Use the cable ringing tool or wire stripper to strip the cable to the proper dimensions Figure 7 Strip SC 22 cable 2 Remove the graphite...

Page 6: ...gure 9 Add hex nut and V gasket to triax cable 2 Install the V clamp as shown in the following figure Carefully seat the V clamp into the V gasket when inserting The sharp edge of the V clamp must spl...

Page 7: ...lamp onto the V clamp Figure 12 Add front clamp 5 Install the gasket clamp insulator and inner clamp as shown in the following figure Figure 13 Add gasket clamp insulator and inner clamp 6 Flare the e...

Page 8: ...ntact Crimp the center contact to the center conductor cable as shown in the following figure Figure 15 Add slotted clamp support insulator and center contact 8 Slide the outer contact over the center...

Page 9: ...ns 071368600 September 2019 9 9 Insert the terminated cable assembly into the rear of the connector body as shown in the following figure 10 Tighten the cable nut with a wrench to 3 39 Newton meters 3...

Page 10: ...ts will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to mains Exercise extreme caution when a...

Page 11: ...r components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component note that selected parts should be purchased only through Keithl...

Reviews: