Section 5: Maintenance
Model 6517B Electrometer User's Manual
5-6
6517B-900-01 Rev. B August 2022
Number
Description
Event
+513
Calibration data lost
EE
+514
DC calibration data lost
EE
+515
Calibration dates lost
+516
Calibration tables lost
EE
+517
Voltage offset lost
EE
+518
Current offset lost
EE
+519
Installed option id lost
EE
+520
Option card not supported
EE
+521
Cal Card Data Error
EE
+522
GPIB communication language lost
EE
+610
Questionable calibration
SE
+611
Questionable temperature
SE
+612
Questionable humidity
SE
+613
Questionable voltage
SE
+614
Questionable current
SE
+615
Questionable ohms
SE
+616
Questionable charge
SE
+617
Questionable test sequence
SE
+618
Resistivity: Current out of limit
EE
+700
Low battery detected
EE
+800
RS-232 framing error detected
EE
+801
RS-232 parity error detected
EE
+802
RS-232 overrun detected
EE
+803
RS-232 break detected
EE
+804
RS-232 noise detected
EE
+805
Invalid system communication
EE
+806
RS-232 settings lost
EE
+807
RS-232 OFLO: Characters Lost
EE
+808
ASCII only with RS-232
EE
+809
Not allowed with RS-232
EE
+850
Invalid test sequence setting
EE
+851
Test sequence running
SE
+860
Interlock violation error
EE
+861
Voltage source limit too low for autoohms
EE
+900
Internal system error
EE
SE = Status event