Section 3: Performance Verification
Model 2231A-30-3 Triple-Channel DC Power Supply Performance Verification
3-22
077102700/ July 2014
DUT voltage selector
switch
AC power source voltage
110
132V
220
253V
9. Enter the voltmeter reading in the Max column for the channel under test. (See Table 22.)
10. Change the AC power source output to the voltage specified in the following table.
DUT voltage selector
switch
AC power source voltage
110
115V
220
230 V
11. Enter the voltmeter reading in the Nom column for the channel under test.
12. Calculate the two values: Nom – Min and Max – Nom. Enter the values in the appropriate
columns.
13. Select the largest of the two calculations from step 12 and enter the value in the Largest column.
14. Turn the DUT output off.
This completes the verification procedure for one channel. If needed, return to step 1 to run through
the procedure for the next channel.