Section 3: Performance Verification
Model 2231A-30-3 Triple-Channel DC Power Supply Performance Verification
3-6
077102700/ July 2014
Instrument test
current Test voltage DUT current Min Measured current Max
Channel 2
0% 15V 0.0000A 0.000000A ------------ 0.01000A
25% 15V 0.7500A 0.73850A ------------ 0.76150A
50% 15V 1.5000A 1.48700A ------------ 1.51300A
75% 15V 2.2500A 2.2355A ------------ 2.2645A
100% 15V 3.0000A 2.9840A ------------ 3.0160A
Instrument test
current Test voltage DUT current Min Measured current Max
Channel 3
0% 3V 0.00000A 0.000000A ----------- 0.01000A
25% 3V 0.7500A 0.73850A ----------- 0.76150A
50% 3V 1.5000A 1.48700A ----------- 1.51300A
75% 3V 2.2500A 2.2355A ----------- 2.2645A
100% 3V 3.0000A 2.9840A ----------- 3.0160A
Table 15: DC current readback accuracy
Instrument test Measured Absolute Maximum
voltage Test current current Current readout difference difference
Channel 1
0% 0A ------------ ------------ ------------ 0.0100A
25% 0.75A ------------ ------------ ------------ 0.0115A
50% 1.5A ------------ ------------ ------------ 0.013A