CH<x>:PRObe:ID:SERnumber? (Query Only)
This query-only command returns the serial number of the probe that is attached to the specified channel. The channel is
specified by x.
Note: For Level 0 and 1 probes, the serial number will be “N/A”.
Syntax
CH<x>:PRObe:ID:SERnumber?
Examples
CH1:PROBE:ID:SERNUMBER?
might return
"B010289"
, indicating that
the serial number of the probe attached to Channel 1 is B010289.
CH<x>:PRObe:ID:TYPe? (Query Only)
This query-only command returns the type of probe that is attached to the specified channel. The channel is specified by x.
Syntax
CH<x>:PRObe:ID:TYPe?
Examples
CH1:PROBE:ID:TYPE?
might return
"TIVP1"
, indicating that a TIVP1
current probe is attached to Channel 1.
CH<x>:PRObe:RESistance? (Query Only)
This query-only command returns the resistance of the probe that is attached to the specified channel. The channel is
specified by x.
Syntax
CH<x>:PRObe:RESistance?
Examples
CH2:PROBE:RESISTANCE?
might return
10.6
, indicating that
the input resistance of the probe attached to Channel 2 is 10 MΩ. This query only
returns valid data when a tip is attached. Otherwise, -1.0000 is returned.
CH<x>:PRObe:SELFCal:State? (Query Only)
This query-only command returns the self-calibration state of RECOMMENDED, RUNNING, or PASSED. The channel is
specified by x.
Syntax
CH<x>:PRObe:SELFCal:State?
Examples
CH1:PRObe:SELFCal:State?
might return
RUNNING
, indicating that the
Channel 1 probe is currently running a self-calibration.
CH<x>:PRObe:SELFCal
This query-only command initiates self-calibration on the probe. The channel is specified by x.
Syntax
CH<x>:PRObe:SELFCal EXECUTE
Examples
CH1:PRObe:SELFCal EXECUTE
runs self-calibration on the Channel 1
probe.
Remote programming
TIVP Series IsoVuTM Measurement System User Manual