![Tektronix CSA7404B Service Manual Download Page 88](http://html1.mh-extra.com/html/tektronix/csa7404b/csa7404b_service-manual_1077968088.webp)
Brief Procedures
4- 12
CSA7404B, TDS7704B, TDS7404B, TDS7254B & TDS7154B Service Manual
H
Pull down the
Horiz/Acq
menu to select
Horizontal/Acquisition
Setup. . .
.
H
Click the
Horizontal
tab in the control window that displays.
H
Click the
Delay Mode
button to turn delay on.
H
Double click the
Horiz Delay
control in the control window to
display the pop-up keypad. Click the keypad buttons to set the
horizontal delay to
1
m
s
and then click the
ENTER
key.
c.
Verify the waveform:
Verify that a rising edge of the waveform is within
a few divisions of center screen.
d.
Adjust the horizontal delay:
Rotate the upper multipurpose knob to
change the horizontal delay setting. Verify that the rising edge shifts
horizontally. Rotate the front-panel horizontal
POSITION
knob. Verify
that this knob has the same effect (it also adjusts delay, but only when
delay mode is on).
e.
Verify the delay toggle function:
H
Rotate the front-panel horizontal
POSITION
knob to center the
falling edge horizontally on the screen.
H
Change the horizontal
SCALE
to
40 ns/div
. The rising edge of the
waveform should remain near the center graticule.
H
Readjust the delay setting to position the rising edge 2 divisions to
the right of the center graticule line.
H
Push the front-panel
DELAY
button several times to toggle delay off
and on and back off again. Verify that the display switches quickly
between two different points in time (the rising edge shifts horizon-
tally on the display).
9.
Remove the test hookup:
Disconnect the BNC cable from the channel input
and the probe compensation output.
Equipment
required
One BNC cable, such as Tektronix part number 012-0076-00
One TCA-BNC adapter
Prerequisites
None
1.
Initialize the instrument:
Push the front-panel
DEFAULT SETUP
button.
2.
Hook up the signal source:
Connect the BNC cable from the probe com-
pensation output to the CH 1 input through a TCA-BNC adapter as shown in
Figure 4--5.
Verify the A (Main) and B
(Delayed) Trigger Systems
Summary of Contents for CSA7404B
Page 4: ......
Page 14: ...Table of Contents x CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 18: ...Service Safety Summary xiv CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 20: ...Environmental Considerations xvi CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 23: ...Specifications ...
Page 24: ......
Page 55: ...Operating Information ...
Page 56: ......
Page 69: ...Theory of Operation ...
Page 70: ......
Page 74: ...Theory of Operation 3 4 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 75: ...Performance Verification ...
Page 76: ......
Page 80: ...Performance Verification 4 4 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 92: ...Brief Procedures 4 16 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 120: ...Performance Tests 4 44 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 215: ...Adjustment Procedures ...
Page 216: ......
Page 218: ...Adjustment Procedures 5 2 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 219: ...Maintenance ...
Page 220: ......
Page 228: ...Maintenance 6 8 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 308: ...Troubleshooting 6 88 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 310: ...Repackaging Instructions 6 90 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 311: ...Options ...
Page 312: ......
Page 319: ...Replaceable Electrical Parts ...
Page 320: ......
Page 322: ...Electrical Parts List 8 2 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 323: ...Diagrams ...
Page 324: ......
Page 327: ...Replaceable Mechanical Parts ...
Page 328: ......
Page 348: ...Mechanical Parts List 10 20 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...