Performance Tests
CSA7404B, TDS7704B, TDS7404B, TDS7254B & TDS7154B Service Manual
4- 89
H
Move the signal to the CH 1 input, change the Trigger Source to
CH 1
, and move the 50
Ω
termination to the CH 4 input.
H
Press
PUSH TO SET 50%
.
H
Set the Vertical SCALE of CH 1, CH 2, CH 3, and CH 4 to
100 mV
.
H
Set the Horizontal SCALE to display 2 to 5 cycles of the signal.
H
Set the generator to output a sine wave at the bandwidth of your
instrument. Set the test signal amplitude for about five divisions on
screen. Now fine adjust the generator output until the
CH 1
Amplitude
readout indicates the amplitude is 500 mV. Readout may
fluctuate around 500 mV.
H
Set the trigger as follows:
TDS7704B: Touch the Source
Line
button.
All other CSA/TDS7000B: Press the front-panel
PUSH TO SET
50%
as necessary to trigger a stable display. At full bandwidth, you
may also want to make small, manual adjustments to the trigger
level. You can use the
Trigger LEVEL
knob to do this.
H
Remove the 2X attenuator.
5.
Confirm the input channels are within limits for channel isolation:
a.
Check — Amplitude of each trace other than CH 1 is 0.67 division or
less (discount trace width). Enter the largest amplitude on the test record.
b.
Move the signal to the CH 2 input connector, change the Trigger
SOURCE to CH 2, and move the 50
Ω
termination to the CH 1 input.
c.
Check — Amplitude of each trace other than CH 2 is 0.67 division or
less (discount trace width). Enter the largest amplitude on the test record.
d.
Move the signal to the CH 3 input connector, change the Trigger
SOURCE to CH 3, and move the 50
Ω
termination to the CH 2 input.
e.
Check — Amplitude of each trace other than CH 3 is 0.67 division or
less (discount trace width). Enter the largest amplitude on the test record.
f.
Move the signal to the CH 4 input connector, change the Trigger
SOURCE to CH 4, and move the 50
Ω
termination to the CH 3 input.
g.
Check — Amplitude of each trace other than CH 4 is 0.67 division or
less (discount trace width). Enter the largest amplitude on the test record.
h.
Select an unchecked Vertical SCALE:
H
Connect the 2X attenuator to CH 1.
Summary of Contents for CSA7404B
Page 4: ......
Page 14: ...Table of Contents x CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 18: ...Service Safety Summary xiv CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 20: ...Environmental Considerations xvi CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 23: ...Specifications ...
Page 24: ......
Page 55: ...Operating Information ...
Page 56: ......
Page 69: ...Theory of Operation ...
Page 70: ......
Page 74: ...Theory of Operation 3 4 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 75: ...Performance Verification ...
Page 76: ......
Page 80: ...Performance Verification 4 4 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 92: ...Brief Procedures 4 16 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 120: ...Performance Tests 4 44 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 215: ...Adjustment Procedures ...
Page 216: ......
Page 218: ...Adjustment Procedures 5 2 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 219: ...Maintenance ...
Page 220: ......
Page 228: ...Maintenance 6 8 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 308: ...Troubleshooting 6 88 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 310: ...Repackaging Instructions 6 90 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 311: ...Options ...
Page 312: ......
Page 319: ...Replaceable Electrical Parts ...
Page 320: ......
Page 322: ...Electrical Parts List 8 2 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 323: ...Diagrams ...
Page 324: ......
Page 327: ...Replaceable Mechanical Parts ...
Page 328: ......
Page 348: ...Mechanical Parts List 10 20 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...