![Tektronix CSA7404B Service Manual Download Page 82](http://html1.mh-extra.com/html/tektronix/csa7404b/csa7404b_service-manual_1077968082.webp)
Brief Procedures
4- 6
CSA7404B, TDS7704B, TDS7404B, TDS7254B & TDS7154B Service Manual
NOTE
. If diagnostic error message 512 is displayed, run signal-path compensa-
tion and then rerun Instrument Diagnostics.
d.
Verify that no failures are found and reported on-screen.
All tests should
pass.
e.
Run the signal-path compensation routine:
H
Pull down the
Utilities
menu and select
Instrument Calibra-
tion. . .
. This displays the instrument calibration control window.
H
If required because the instrument is in service mode, select the
Signal Path
button under Calibration Area.
H
Click the
Calibrate
button to start the routine.
f.
Wait:
Signal-path compensation may take five to fifteen minutes to run.
g.
Confirm signal-path compensation returns passed status:
Verify that the
word
Pass
appears in the instrument calibration control window.
2.
Return to regular service:
Click the
Close
button to exit the instrument
calibration control window.
Functional Tests
The purpose of these procedures is to confirm that the instrument functions
properly. The only equipment required is a P7240 probe (P7260 probe with
TDS7704B), a probe calibration and deskew fixture, a BNC cable, BNC-to-SMA
adapter or TCA-BNC adapter, and, to check the file system, a 3.5 inch,
1.44 Mbyte, formatted floppy disk.
STOP
. These procedures verify functions; that is, they verify that the instrument
features operate. They do not verify that they operate within limits.
Therefore, when the instructions in the functional tests that follow call for you to
verify that a signal appears on-screen “that is about five divisions in amplitude”
or “has a period of about six horizontal divisions,” etc., do NOT interpret the
quantities given as limits. Operation within limits is checked in
Performance
Tests
, which begin on page 4--17.
Summary of Contents for CSA7404B
Page 4: ......
Page 14: ...Table of Contents x CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 18: ...Service Safety Summary xiv CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 20: ...Environmental Considerations xvi CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 23: ...Specifications ...
Page 24: ......
Page 55: ...Operating Information ...
Page 56: ......
Page 69: ...Theory of Operation ...
Page 70: ......
Page 74: ...Theory of Operation 3 4 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 75: ...Performance Verification ...
Page 76: ......
Page 80: ...Performance Verification 4 4 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 92: ...Brief Procedures 4 16 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 120: ...Performance Tests 4 44 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 215: ...Adjustment Procedures ...
Page 216: ......
Page 218: ...Adjustment Procedures 5 2 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 219: ...Maintenance ...
Page 220: ......
Page 228: ...Maintenance 6 8 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 308: ...Troubleshooting 6 88 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 310: ...Repackaging Instructions 6 90 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 311: ...Options ...
Page 312: ......
Page 319: ...Replaceable Electrical Parts ...
Page 320: ......
Page 322: ...Electrical Parts List 8 2 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...
Page 323: ...Diagrams ...
Page 324: ......
Page 327: ...Replaceable Mechanical Parts ...
Page 328: ......
Page 348: ...Mechanical Parts List 10 20 CSA7404B TDS7704B TDS7404B TDS7254B TDS7154B Service Manual ...