Operating Basics
Set up load impedance
The output impedance of the instrument is 50
Ω
. If you connect a load other than
50
Ω
, the displayed Amplitude, Offset, and High/Low values will be different
from the output voltage. To make the displayed values the same as the output
voltage, you need to set the load impedance. The load impedance is applied to
the amplitude, offset, and high/low level settings. To set the load impedance,
use the Output Menu.
1.
From the top menu, push
the Output Menu bezel
button.
2.
Push Load Impedance
to display the Load
Impedance submenu.
3.
To adjust the load
impedance, select Load.
4.
You can set the load
impedance to any value
from 1
Ω
to 10 k
Ω
.
5.
When the load
impedance is set to
other than 50
Ω
, the set
value is displayed in the
output status.
NOTE.
When dBm is speci
fi
ed for output amplitude units, the amplitude units
setting is automatically changed to Vpp if you select high impedance.
AFG2021 Quick Start User Manual