Calibration—4 8 5 /R 4 8 5 Service
d.
INTERACTION with 5 ns timing, repeat steps 26 and
27 if necessary.
31. ADJUST 1 ns T IM IN G
a. Set TIM E/D IV to 1 ns.
b.
Adjust 1 ns (R1228), (located in back of the lower
relay on the Sweep board) for exactly 1 to 9 timing (four
2 ns markers over the center 8 divisions).
c.
If exact timing cannot be achieved, a compromise
adjustment may be done.
32. CHECK 1, 2, and 5 ns SWEEP L IN E A R IT Y
a.
Less than 0.1 division of error with timing adjusted
on (1 ns through 20 ns). With time marks exactly aligned at
1 and 9 the remaining marks must be no greater than
0.1 div away from their respective graticule lines. The
incremental error must not exceed 0.1 div over any one
division.
33. CHECK B SWEEP T IM IN G ACCURACY
a.
Check timing over center 8 graticule divisions; posi
tion one time mark to 1 and read error at 9.
b. 1 ns to 20 ns;—within 0.16 div (2%).
c. 50 ns to 0.1 s;—within 0.1 div (1.25%).
d. 0.2 s to 0.5 s;—within 0.16 div (2%).
34. CHECK A SWEEP T IM IN G ACCURACY
a. 1 ns to 20 ns;—within 0.16 div (2%).
b. 50 ns to 0.1 s;—within 0.1 div (1.25%).
c. 0.2 s to 0.5 s;—within 0.16 div (2%).
35. CHECK A U TO R E P E TITIO N RATE
a.
Set A TIM E/DIV to 50
ms
; SWEEP MODE to AUTO
TRIG.
b.
Check that sweep triggers on 50 ms markers and will
not trigger on 0.1 s markers.
36. CHECK D E LA Y T IM E A C C U R A C Y , 10 ns/
div to 0.5 s/div
a.
Use time marks that give 1 to 5 markers/div on the A
display. Use B TIM E/D IV for 1 ns for A TIM E/D IV settings
of 10 ns through 0.1 Ms. For A TIM E/D IV setting of 0 .2
ms
and slower maintain an A:B ratio of 100:1.
b.
Set the DTP dial exactly to the reference position
(1.0 except at the fastest speeds). Use the POSITION
(HORIZ) control to move the marker on the B display to a
reference position.
c.
Rotate the DTP to 9.0 and then slightly from 9.0 to
place the B display marker to the same screen reference
position. Read the error (from 9.00) in the minor dial
divisions on the DTP dial.
TABLE 5-4
A Time/Div
Reference
DTP
Position
Maximum DTP Dial Error at
Setting near 9.0 (minor
divisions)
10 ns
4.0
6
20 ns
2.0
7
50 ns to 1 ms
1.0
4
2 ms to 0.5 s
1.0
8
37. CHECK
IN C R E M E N T A L
D E L A Y
ACCU
RACY 10, 20, and 50 ns/div
NOTE
Check the erro r at each turn against the requirements
given in the chart a t the end o f this step. Detailed
procedure is given fo r the 10 ns/div.
a.
A and B TIM E/D IV of 10ns and In s ; HORIZ
DISPLAY to B; A TRIGGER HOLDOFF to NORM; time
marks of 10 ns.
b.
Set the DTP dial to exactly 4.0. Adjust the POSI
TION (HORIZ) for marker leading edge at midscreen. Turn
the DTP dial CW one turn at a time and bring a marker to
the exact midscreen position.
c.
Check that the dial errors at 5.0, 6.0, and 7.0 are
within the allowable limits (at 10 ns/div the change in error
over any one turn is not to exceed 2 minor dial divisions,
and the total error at any turn is not to exceed 3 minor dial
divisions).
5-14
Summary of Contents for 485
Page 3: ...485 R485 rviice T h 5 0 Oscil oscope ...
Page 74: ...Rackmounting 485 R485 Service 6 6 ...
Page 128: ...485 R485 Service MAR 1979 5 m V B A L ADJ ...
Page 152: ... 0 iT s i ...
Page 164: ...2S2 n eKn s e s 5 4 85 Os c illo sc o p e REV F A U S 1 9 7 7 I 1 9 3 5B POWER SUPPLY 773 ...
Page 178: ......
Page 179: ...485 R485 OSCILLOSCOPE REV D MAR 1977 ...
Page 184: ...REV C MAR 1976 485 R485 OSCILLOSCOPE ...
Page 197: ......