![Tektronix 485 Instruction Manual Download Page 47](http://html1.mh-extra.com/html/tektronix/485/485_instruction-manual_1078906047.webp)
TABLE 5-1
TEST EQUIPMENT (cont)
Calibration-485/R485 Service
Description
Minimum Specifications
Usage
Examples of Applicable
Test Equipment
Square-Wave
Generator
Frequency Range, 100 Hz to
100 kHz; Amplitude variable
0.5 to 12 V.
Adjust: 50 12 high frequency
transient response; 5012 low
frequency transient response;
1 M12 10 kHz transient re
sponse; 1 M12 input attenu
ator compensations; Ext Trig
indent delay match; A Ext
Trig response.
T E K T R O N IX
Type
106
Square-Wave Generator.
2X Attenuator1
(2 required)
Impedance, 50 JT2; accuracy,
2%; BNC connectors.
Used throughout the pro
cedure for signal attenuation.
TEKTRONIX
Part
No.
011-0069-02.
5X Attenuator1
Impedance, 50 12; accuracy,
2%; BNC connectors.
Used throughout the pro
cedure for signal attenuation.
TEKTRONIX
Part
No.
011-0060-02.
10X Attenuator
Impedance, 50 12; accuracy,
2%; BNC connector.
Used throughout the pro
cedure for signal attenuation.
TEKTRONIX
Part
No.
011-0059-02.
T Connector
Connector, BNC.
Ext Trig checks.
T E K T R O N IX
Part
No.
103-0030-00.
Termination (2
required)
Impedance, 50 12; accuracy,
2%; connectors, BNC.
CMR check.
T E K T R O N IX
Part
No.
011-0049-01.
Cable (2 required)
Impedance, 50 12; type RG-
58/U; length, 42 inch; con
nectors, BNC.
Used throughout the pro
cedure for signal intercon
nection.
TEKTRONIX
Part
No.
012-0057-01.
GR thru-line
termination
Impedance, 50 12; accuracy,
2%; connectors, GR874 to
BNC male.
1 M12 bandwidth check.
T E K T R O N IX
Part
No.
017-0083-00.
Adapter
GR874 to BNC male.
50 12
vertical
bandwidth
check.
T E K T R O N IX
Part
No.
017-0064-00.
Screwdriver
Three-Inch shaft, 3/32 inch
bit.
Used throughout the pro
cedure to adjust variable resis
tors.
Xcelite R-3323.
Low Capacitance
Screwdriver1
1 1/2 inch shaft.
Used throughout the pro
cedure to adjust variable cap
acitors.
T E K T R O N IX
Part
No.
003-0000-00.
Nylon Tuning
Tool1
Fits 5/64 inch (ID) hex cores.
X-Y phasing.
H a n d l e
a n d
i n s e r t
T E K T R O N IX
Part
No.
0 0 3 - 0 3 0 7 - 0 0
a n d
003-0310-00.
R EV. B, M AR. 1977
5-3
Summary of Contents for 485
Page 3: ...485 R485 rviice T h 5 0 Oscil oscope ...
Page 74: ...Rackmounting 485 R485 Service 6 6 ...