Table 4 (cont)
Characteristics
Performance Requirements
Supplemental Inform ation
TRIGGER SYSTEM (cont)
Trigger View
Deflection Factor
EXT
,lL
- --
100 mV per division ±40%.
E X T - 10
1 V per division ±40%.
Centering o f Trigger Point
Within 1 division o f center screen.
Bandwidth
To at least 80 MHz.
4-division reference signal from a 2 5-0
source; centered vertically.
Delay Difference
3 ns ±2 ns.
5-division signal w ith 5-ns rise time or
less from 2 5-0 source, centered
vertically; equal cable length from
signal source to vertical channel and
external trigger inputs, terminated in
50 O at each input.
2335 Operators
73