Figure 20. High-resoiution phase difference.
shows the same signals illustrated in Figure 19, but
magnifying the displays results in a horizontal difference
o f 6 divisions between the tw o signals.
Substituting the given values in the phase difference
formula:
Phase Difference = 6 div x 4.5° /div = 27°
Amplitude Comparison
In some applications it may be necessary to establish
a set of deflection factors other than those indicated by
the VOLTS/D1V switch settings. This is useful for com
paring unknown signals to a reference signal o f known
amplitude. To accomplish this, a reference signal o f known
amplitude is first set to an exact number of vertical divisions
by adjusting the VO LTS/DIV switch and VAR control.
2335 Operators
49