12.
Calculate the time difference between the Channel 1
and Channel 2 pulses using the same form ula listed in the
“ Time D uration" measurement procedure.
EXAMPLE: With the A SEC/DIV switch set to 50
fis,
the B D ELAY TIME POSITION dial reading fo r the
reference pulse (Channel 1) is 2.60 and the dial reading
fo r the comparison pulse (Channel
2)
is 7.10.
Substituting the given values in the time difference
formula:
Time Difference = (7.10 — 2.60) (50 jus) = 225 jus