Adjustment Procedure—2205 Service
f. ADJUST-HF Comp (C241 and R241) for best
flat top on the front corner.
g. Repeat parts
e
and f until on further improve
ment is noted.
h. Set the CH 1 VOLTS/DIV switch to 5 mV,
i. Set the generator to produce a 5-division
display.
j. Check for aberrations of 5% (0.3 division) or
less.
k. Repeat part j for each CH 1 VOLTS/DIV switch
settings from 5 mV through 0.5 V. Adjust the gen
erator output and add or remove the 10X attenuator
as necessary to maintain a 5-division display at
each VOLTS/DIV switch setting.
l
. Move the cable from the CH 1
or
X input con
nector to the CH 2 OR Y input connector. Set the
Vertical MODE switch to CH 2.
m. Repeat part k for Channel 2.
n. Disconnect the test equipment from the
instrument.
f. Repeat parts c through e for all CH 1 VOLTS/
DIV switch settings, up to the output-voltage upper
limit of the sine-wave generator being used.
g. Move the cable from the CH 1
o r
X input con
nector to the CH 2
o r y
input connector. Set the
Vertical MODE switch to CH 2.
h. Repeat parts c through e for all CH 2 VOLTS/
DIV switch settings, up to the output-voltage upper
limit of the sine-wave generator being used.
14. Check Channel Isolation
a. Set:
Channel 1 AC-GND-DC
GND
b. Set the generator to produce a 20-MHz,
5-division display.
c. Set CH 2 VOLTS/DIV switch to 0.5 V for a
10-division display.
d. Set Vertical MODE switch to CH 1 and ALT.
NOTE
e.
CHECK—The display amplitude is less than 0.1
division. :
Install the instrument cabinet for the re
maining
vertical
checks
and
allow
a
20-minute warm-up period before continuing
with the "Adjustment Procedure". See the
"
Cabinet“ remove and replace instructions
located in the "Maintenance" section of the
manual.
13. Check Bandwidth
a. Set:
f.
Move the test-signal cable from the CH 2
OR
Y
input connector to the CH 1
OR
X input connector.
g- Set:
Vertical MODE
CH 2
Channel 1 AC-GND-DC
DC
Channel 2 AC-GND-DC
GND
h.
CHECK—The display amplitude is less than 0.1
division.
VOLTS/DIV (both)
Vertical MODE
SEC/DIV
Trigger SOURCE
5 mV
CH 1
10 ms
VERT MODE
i. Disconnect the test equipment from the
instrument.
15. Check Comm on-M ode Rejection Ratio
b. Connect the leveled sine-wave generator out
put via a 5 0 -0 coaxial cable and a 5 0 -0 termination
to the CH 1
or
X input connector.
c. Set the generator to produce a 5 0 -kHz, 6-
division display.
d. Increase the sine-wave frequency until a
4.2-diviston display is obtained.
e. CHECK—the frequency is greater than 20 MHz.
a. Set:
VOLTS/DIV (both)
10 mV
Channel 2 AC-GND-DC
DC
Vertical MODE
BOTH, NORM.
and ALT
b. Connect the leveled sine-wave generator out
put via a 5 0 -0 coaxial cable, a 5 0 -0 termination,
and a dual-input coupler to the CH 1
or
X and CH 2
OR Y input connectors.
f
i
I
5—10
Summary of Contents for 2205
Page 1: ...I f O M r T 7 i ESC 3234 95 01 service Manual Tektronix 2205 Oscilloscope 070 6716 00 ...
Page 2: ... ...
Page 14: ...6716 01 ...
Page 22: ...Specification 2205 Service 6718 03 1 8 Figure 1 2 Instrument dimensional drawing ...
Page 32: ......
Page 50: ...r i ...
Page 82: ......
Page 100: ...i ...
Page 132: ......
Page 142: ...PIG 3 ACCESSORIES QO t Q 2205 Service ...
Page 148: ...i 4 ...
Page 150: ......
Page 154: ...s ...
Page 156: ...i ...
Page 158: ......
Page 164: ......
Page 166: ...f ...
Page 168: ......
Page 170: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 179: ...2205 Service 6716 18 Figure 9 4 Block diagram ...
Page 187: ......
Page 191: ...L y____ K____ j _____J i y_____E____j _____E _____D____j _____C ____ B____j _____fl 220 J ...
Page 192: ...WAVEFORMS FOR DIAGRAM 2 AC Waveforms VERTICAL MODE BOTH andCHOP ...
Page 196: ...FIG 9 10 2205 Service c i i ...
Page 203: ......
Page 206: ...Figure 9 13 A2 Timebase Attenuator board adjustment locations ...
Page 207: ...C392 2205 Service R84 CH 2 INVERT BAL ...
Page 208: ......
Page 209: ...2205 Service ...
Page 210: ... o m 2205 S e r v i c e ...