Basic Measurements
1760 Series Component Waveform/Vector Monitor User Manual
4-5
Making Waveform Measurements
Horizontal Sync Amplitude
The vertical scale at sync tip (40 IRE NTSC or --0.3 V PAL)
provides
±
2 and
±
4 IRE (
±
2 and
±
4 mV) for measuring sync
amplitude. Align the blanking level of a centered two
-
line
waveform display on the graticule baseline to place the sync
pulse directly over this structure. Measure sync amplitude
directly from the graticule, or use the voltage cursors.
Peak White
With the blanking signal level on the graticule baseline, the
amplitude of the 100% white reference of a color bar should
be on the 100 IRE (1 V) graticule line. With a
2 LINE
sweep
rate, use the
HORIZ POS
control to place the white reference
directly over the amplitude scale at graticule center. The
vertical markings there correspond to 2 IRE/division (10 mV/
division for PAL). Measure peak white directly from the
graticule or use the voltage cursors.
K-Factor Measurements
The boxed area in the upper left corner of the graticule is
scaled for 2% and 4% K--Factor ratings for precise tilt mea-
surements. This structure is designed to work with an 8
µ
s
half-amplitude duration (HAD) bar. The bar tilt measure-
ment structure can also be used to measure pulse-to-bar ratio.
Calibrated
X5 GAIN
can be used while making these measure-
ments; it increases resolution to 0.4% and 0.8%.
To use the NTSC vertical scale to make line-time distortion
and pulse-to-bar ratio measurements, position the signal
blanking level to the graticule baseline. Next, position the
center of the composite test signal bar to the center of the 2T
bar measurement box and check to see if insertion gain is
unity. If it is not, adjust the
VARIABLE GAIN
(enabled from the
GAIN MENU
) for exactly 100 IRE (700 mV PAL) of signal
amplitude from baseline to the middle of the white bar.
Summary of Contents for 1760 series
Page 4: ......
Page 8: ...General Safety Summary 1760 Series Component Waveform Vector Monitor User Manual iv ...
Page 19: ...Introduction ...
Page 20: ......
Page 24: ...Product Description 1 4 1760 Series Component Waveform Vector Monitor User Manual ...
Page 34: ...Installation 1760 Series Component Waveform Vector Monitor User Manual 1 14 ...
Page 36: ...Instrument Configuration 1760 Series Component Waveform Vector Monitor User Manual 1 16 ...
Page 37: ...Getting Started ...
Page 38: ......
Page 67: ...Operation Basics ...
Page 68: ......
Page 96: ...Using the Menus 1760 Series Component Waveform Vector Monitor User Manual 3 28 ...
Page 98: ...Remote Operation 1760 Series Component Waveform Vector Monitor User Manual 3 30 ...
Page 99: ...Making Measurements ...
Page 100: ......
Page 124: ...Basic Measurements 1760 Series Component Waveform Vector Monitor User Manual 4 24 ...
Page 127: ...Appendices ...
Page 128: ......
Page 152: ...Appendix B Remote Connectors 1760 Series Component Waveform Vector Monitor User Manual B 6 ...
Page 160: ...Appendix D Software Version 1760 Series Component Waveform Vector Monitor User Manual D 2 ...
Page 161: ...Glossary ...
Page 162: ......
Page 173: ...Index ...
Page 174: ......
Page 180: ...Index 1760 Series Component Waveform Vector Monitor User Manual I 6 ...