Clear and Sanitize Procedures

Table 1: Volatile Memory Devices (cont.)

Type and
minimum size

Function

User
modi

fi

able

1

Data input
method

Location

To clear

To sanitize

2

SDRAM 16 M X 16

CPU RAM

No

CPU access

Main board

Remove the power
source from the
instrument for at least 20
seconds

Remove the Main
board; see the product
service manual for
removal instructions.
Store the removed
board in a secure
area or destroy it.
Contact Tektronix for
board replacement
and adjustment

SRAM 256 K X 16

PLD RAM

No

PLD access

Main board

Remove the power
source from the
instrument for at least 20
seconds

Remove the Main
board; see the product
service manual for
removal instructions.
Store the removed
board in a secure
area or destroy it.
Contact Tektronix for
board replacement
and adjustment

1

During normal instrument operation.

2

Before you remove any module or board for the purpose of sanitizing the instrument, please contact your local Tektronix Customer Service Center to check on

the availability and cost of replacement parts.

Table 2: Nonvolatile Memory Devices

Type and
minimum size

Function

User
modi

fi

able

1

Data input
method

Location

To clear

To sanitize

2

Serial EEPROMs
256 X 8

Stores Composite
calibration
coef

fi

cients

No

Programmed
during
calibration

Composite boards

Recalibrate
Composite

Remove the Composite
boards; see the product
service manual for removal
instructions. Store the
removed board in a
secure area or destroy
it. Contact Tektronix for
board replacement and
adjustment

Flash Memory
512 M X 16

Loads FPGAs on
power up
Contains
instrument SW,
network access
parameters, and
user-de

fi

ned

presets

Yes

Programmed
by software
during software
upgrade

Main board

(See page 4,

Clear Presets
Procedure

.)(See

page 5,

Clear IP

Address Fields
Procedure

.)

Remove the Main board;
see the product service
manual for removal
instructions. Store the
removed board in a
secure area or destroy
it. Contact Tektronix for
board replacement and
adjustment

PLD 8 Kb

Display controller

No

Programmed by

fl

ash memory

during power
up

Main board

None

Remove the Main board;
see the product service
manual for removal
instructions. Store the
removed board in a
secure area or destroy
it. Contact Tektronix for
board replacement and
adjustment

1741C Declassi

fi

cation and Security

3

Summary of Contents for 1741C

Page 1: ...1741C Analog Dual Standard Waveform Monitor Declassification and Security ZZZ Instructions xx www tektronix com 077 0153 00...

Page 2: ...and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tek...

Page 3: ...Sanitize Procedures 1 Memory Devices 1 Data Export Devices 6 Troubleshooting 7 How to Clear or Sanitize a Nonfunctional Instrument 7 How to Recover from Clearing or Removing the Instrument s Memory 7...

Page 4: ...Table of Contents ii 1741C Declassification and Security...

Page 5: ...t is not functioning Products This document covers the Tektronix 1741C Analog Dual Standard Waveform Monitor Related Documents Document Part number Purpose 1741C Analog Dual Standard Waveform Monitor...

Page 6: ...mode does not qualify as powering off For these products you will need to either press a rear panel OFF switch or remove the power source from the instrument Remove This is a physical means to clear...

Page 7: ...in board Remove the power source from the instrument for at least 20 seconds Remove the Main board see the product service manual for removal instructions Store the removed board in a secure area or d...

Page 8: ...d see the product service manual for removal instructions Store the removed board in a secure area or destroy it Contact Tektronix for board replacement and adjustment FPGA 1 9 Mb Rasterizer No Progra...

Page 9: ...Nonvolatile Memory Devices Type and minimum size Function User modifiable1 Data input method Location To clear To sanitize 2 Serial EEPROMs 256 X 8 Stores Composite calibration coefficients No Program...

Page 10: ...placement and adjustment Serial EEPROM MAC address No Set at factory Main board None Remove the Main board see the product service manual for removal instructions Store the removed board in a secure a...

Page 11: ...hlighted Then press SEL to exit the log display 4 Press the CONFIG button to exit the configuration menu Clear IP Address Fields Procedure 1 Press the CONFIG button Then select Network Settings 2 Set...

Page 12: ...r of instrument See page 6 Disable Ethernet Access Procedure USB Data storage Yes Standard USB protocol Front of instrument None 1 During normal instrument operation Disable Ethernet Access Procedure...

Page 13: ...the instrument to Tektronix for installation of a new Main board This procedure does not clear calibration constants stored on the Audio Eye and Composite boards How to Recover from Clearing or Remov...

Reviews: