background image

Clear and Sanitize Procedures

Table 2: Nonvolatile Memory Devices (cont.)

Type and
minimum size

Function

User
modi

fi

able

1

Data input
method

Location

To clear

To sanitize

2

Real Time Clock

Stores time set by
user

Yes

UI

Main board

Set to GMT

Set to GMT

Serial EPROM

Diagnostic logs,
Main board Pix
Mon calibration,
and factory only
Looping diagnostic
mode control

No

Programmed by
software during
calibration and
diagnostics

Main board

(See page 5,

Clear

Diagnostic Log
Procedure

.)

Recalibrate Pix mon
to overwrite existing
values.

Remove the Main board;
see the product service
manual for removal
instructions. Store the
removed board in a
secure area or destroy
it. Contact Tektronix for
board replacement and
adjustment

Serial Flash

Future expansion

No

Tested on
power up
Not used

Main board

None

Remove the Main board;
see the product service
manual for removal
instructions. Store the
removed board in a
secure area or destroy
it. Contact Tektronix for
board replacement and
adjustment

Serial EEPROM

MAC address

No

Set at factory

Main board

None

Remove the Main board;
see the product service
manual for removal
instructions. Store the
removed board in a
secure area or destroy
it. Contact Tektronix for
board replacement and
adjustment

1

During normal instrument operation.

2

Before you remove any module or board for the purpose of sanitizing the instrument, please contact your local Tektronix Customer Service Center to check on

the availability and cost of replacement parts.

Clear Presets Procedure

1.

Press and hold the

PRESET

button to display the preset con

fi

guration menu.

Then select

Recall Preset

>

Factory Preset

>

SEL

. This restores the presets

to the factory state.

2.

Press and hold the

PRESET

button. Then select

Save Preset

>

Select Group

A

. For all presets that are not empty “<e>”, select

Save

and then press

SEL

to overwrite that preset with the factory settings.

3.

Check each group and overwrite presets that are not empty as described in
the previous step.

4.

Navigate to the top level menu and select

Rename Preset

. Rename all named

groups and presets with the label “Default”.

4

1741C Declassi

fi

cation and Security

Summary of Contents for 1741C

Page 1: ...1741C Analog Dual Standard Waveform Monitor Declassification and Security ZZZ Instructions xx www tektronix com 077 0153 00...

Page 2: ...and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tek...

Page 3: ...Sanitize Procedures 1 Memory Devices 1 Data Export Devices 6 Troubleshooting 7 How to Clear or Sanitize a Nonfunctional Instrument 7 How to Recover from Clearing or Removing the Instrument s Memory 7...

Page 4: ...Table of Contents ii 1741C Declassification and Security...

Page 5: ...t is not functioning Products This document covers the Tektronix 1741C Analog Dual Standard Waveform Monitor Related Documents Document Part number Purpose 1741C Analog Dual Standard Waveform Monitor...

Page 6: ...mode does not qualify as powering off For these products you will need to either press a rear panel OFF switch or remove the power source from the instrument Remove This is a physical means to clear...

Page 7: ...in board Remove the power source from the instrument for at least 20 seconds Remove the Main board see the product service manual for removal instructions Store the removed board in a secure area or d...

Page 8: ...d see the product service manual for removal instructions Store the removed board in a secure area or destroy it Contact Tektronix for board replacement and adjustment FPGA 1 9 Mb Rasterizer No Progra...

Page 9: ...Nonvolatile Memory Devices Type and minimum size Function User modifiable1 Data input method Location To clear To sanitize 2 Serial EEPROMs 256 X 8 Stores Composite calibration coefficients No Program...

Page 10: ...placement and adjustment Serial EEPROM MAC address No Set at factory Main board None Remove the Main board see the product service manual for removal instructions Store the removed board in a secure a...

Page 11: ...hlighted Then press SEL to exit the log display 4 Press the CONFIG button to exit the configuration menu Clear IP Address Fields Procedure 1 Press the CONFIG button Then select Network Settings 2 Set...

Page 12: ...r of instrument See page 6 Disable Ethernet Access Procedure USB Data storage Yes Standard USB protocol Front of instrument None 1 During normal instrument operation Disable Ethernet Access Procedure...

Page 13: ...the instrument to Tektronix for installation of a new Main board This procedure does not clear calibration constants stored on the Audio Eye and Composite boards How to Recover from Clearing or Remov...

Reviews: