Clear and Sanitize Procedures

Data Export Devices

The following table lists the data export devices. Detailed procedures to disable
these devices, if any, are shown following the table.

Table 3: Data Export Devices

Type and
minimum
size

Function

User
modi

fi

able

1

Data input method

Location

To disable

Ethernet

Communications

Yes

Standard Ethernet
protocol

Rear of instrument

(See page 6,

Disable Ethernet Access

Procedure

.)

USB

Data storage

Yes

Standard USB protocol

Front of instrument

None

1

During normal instrument operation.

Disable Ethernet Access

Procedure

1.

Press the

CONFIG

button. Then select

Network Settings

.

2.

Navigate to the

Web Enable

and set it to

Off

.

3.

Press the

CONFIG

button to exit the con

fi

guration menu.

Enable Ethernet Access

Procedure

To enable Ethernet access, use the same procedures you would use to disable
these devices, but select

On

to enable each device.

6

1741C Declassi

fi

cation and Security

Summary of Contents for 1741C

Page 1: ...1741C Analog Dual Standard Waveform Monitor Declassification and Security ZZZ Instructions xx www tektronix com 077 0153 00...

Page 2: ...and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tek...

Page 3: ...Sanitize Procedures 1 Memory Devices 1 Data Export Devices 6 Troubleshooting 7 How to Clear or Sanitize a Nonfunctional Instrument 7 How to Recover from Clearing or Removing the Instrument s Memory 7...

Page 4: ...Table of Contents ii 1741C Declassification and Security...

Page 5: ...t is not functioning Products This document covers the Tektronix 1741C Analog Dual Standard Waveform Monitor Related Documents Document Part number Purpose 1741C Analog Dual Standard Waveform Monitor...

Page 6: ...mode does not qualify as powering off For these products you will need to either press a rear panel OFF switch or remove the power source from the instrument Remove This is a physical means to clear...

Page 7: ...in board Remove the power source from the instrument for at least 20 seconds Remove the Main board see the product service manual for removal instructions Store the removed board in a secure area or d...

Page 8: ...d see the product service manual for removal instructions Store the removed board in a secure area or destroy it Contact Tektronix for board replacement and adjustment FPGA 1 9 Mb Rasterizer No Progra...

Page 9: ...Nonvolatile Memory Devices Type and minimum size Function User modifiable1 Data input method Location To clear To sanitize 2 Serial EEPROMs 256 X 8 Stores Composite calibration coefficients No Program...

Page 10: ...placement and adjustment Serial EEPROM MAC address No Set at factory Main board None Remove the Main board see the product service manual for removal instructions Store the removed board in a secure a...

Page 11: ...hlighted Then press SEL to exit the log display 4 Press the CONFIG button to exit the configuration menu Clear IP Address Fields Procedure 1 Press the CONFIG button Then select Network Settings 2 Set...

Page 12: ...r of instrument See page 6 Disable Ethernet Access Procedure USB Data storage Yes Standard USB protocol Front of instrument None 1 During normal instrument operation Disable Ethernet Access Procedure...

Page 13: ...the instrument to Tektronix for installation of a new Main board This procedure does not clear calibration constants stored on the Audio Eye and Composite boards How to Recover from Clearing or Remov...

Reviews: