background image

1741C

Analog Dual-Standard Waveform Monitor

Declassi

fi

cation and Security

ZZZ

Instructions

xx

www.tektronix.com

077-0153-00

Summary of Contents for 1741C

Page 1: ...1741C Analog Dual Standard Waveform Monitor Declassification and Security ZZZ Instructions xx www tektronix com 077 0153 00...

Page 2: ...and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tek...

Page 3: ...Sanitize Procedures 1 Memory Devices 1 Data Export Devices 6 Troubleshooting 7 How to Clear or Sanitize a Nonfunctional Instrument 7 How to Recover from Clearing or Removing the Instrument s Memory 7...

Page 4: ...Table of Contents ii 1741C Declassification and Security...

Page 5: ...t is not functioning Products This document covers the Tektronix 1741C Analog Dual Standard Waveform Monitor Related Documents Document Part number Purpose 1741C Analog Dual Standard Waveform Monitor...

Page 6: ...mode does not qualify as powering off For these products you will need to either press a rear panel OFF switch or remove the power source from the instrument Remove This is a physical means to clear...

Page 7: ...in board Remove the power source from the instrument for at least 20 seconds Remove the Main board see the product service manual for removal instructions Store the removed board in a secure area or d...

Page 8: ...d see the product service manual for removal instructions Store the removed board in a secure area or destroy it Contact Tektronix for board replacement and adjustment FPGA 1 9 Mb Rasterizer No Progra...

Page 9: ...Nonvolatile Memory Devices Type and minimum size Function User modifiable1 Data input method Location To clear To sanitize 2 Serial EEPROMs 256 X 8 Stores Composite calibration coefficients No Program...

Page 10: ...placement and adjustment Serial EEPROM MAC address No Set at factory Main board None Remove the Main board see the product service manual for removal instructions Store the removed board in a secure a...

Page 11: ...hlighted Then press SEL to exit the log display 4 Press the CONFIG button to exit the configuration menu Clear IP Address Fields Procedure 1 Press the CONFIG button Then select Network Settings 2 Set...

Page 12: ...r of instrument See page 6 Disable Ethernet Access Procedure USB Data storage Yes Standard USB protocol Front of instrument None 1 During normal instrument operation Disable Ethernet Access Procedure...

Page 13: ...the instrument to Tektronix for installation of a new Main board This procedure does not clear calibration constants stored on the Audio Eye and Composite boards How to Recover from Clearing or Remov...

Reviews: