4.3.7
Steps to Configure the Parameters
1.
Pipe outer diameter.
2.
Pipe wall thickness.
3.
Pipe materials.
4.
Linear material and its sound speed and thickness shows half sentence what any
linear.
5.
For non-standard liquids, the sound speed of the liquid is also needed.
6.
Generally, the Standard M1 clamp-on transducers is selected.
7.
Generally, Transducer mounting methods are the V-method or Z-method is the
common option.
8.
Check the space displayed on M25 and install transducer accordingly.
For standard pipe materials and standard liquids, the following detailed step-by-step
setup is recommended.
1.
Press keys
MENU
diameter, and then press 11
ENT
to enter M11 window to input the
digits for the pipe outer key.
2.
Press key
▼
/- to enter M12 window to input the digits for the pipe outer diameter
and then press
ENT
key.
3.
Press key
▼
/- (down) to enter M14 window, and press mode. Use keys
▲
/+ (Up)
and
▼
/- (down) to scroll up and down to the intended pipe material, and then
press
ENT
.
4.
Press key
▼
/- (down) to enter M16 window, press
ENT
key to enter the option selection
ENT
key to
enter the option selection mode, use keys
▲
/+ (Up) and
▼
/- (down) to scroll up
and down to the liner material, and then press key. Select
“
No Liner”, if there is no
liner.
5.
Press key
▼
/- (down) to enter M20 window, press
ENT
key to enter the option
selection mode, use keys
▲
/+ (Up) and
▼
/- (down) to scroll up and down to the
proper liquid, and then press
ENT
key
6.
Press key
▼
/- (down) to enter M23 window, press key to enter the option selection
mode,
ENT
use keys
▲
/+ (Up) and
▼
/- (down) to scroll up and down to the proper
transducer type, and then press
ENT
key.
7.
Press key
▼
/- (down) to enter M24 window, press use keys
▲
/+ (Up) and
▼
/-
(down) to scroll up and down to the proper transducer mounting method, and then
press
ENT
key
8.
Press key
▼
/- (down) to enter M25 window to install the transducers on the pipe,
and then press key to go to M01 for the results.
The first-time users may need some time to get familiar with the operation. the user-
friendly interface of the device makes the operation quite smooth and simple. You will
soon find that it is very quick to configure the instrument with very little key pressing,
since the interface allows the user to go to the desired operation directly without any
extra steps.