Troubleshooting

Always check the unit and accessories before use to prevent damage and defects.
1.  Make sure the batteries have sufficient charge and are not corroded.
2.  Make sure the cables fit tightly into the connection sockets of the device.
 

If you cannot remedy the defects  as described below, contact Customer Service.

Problem

Possible Cause

Solutions

The device does not 

turn on

No battery or bad battery

Replace batteries

The device turns on

and then off again

Battery not inserted properly

Battery life expired

Insert batteries again or 

replace batteries

The device turns on,

but does not generate 

electric pulses

Intensity is too low

Increase the intensity

Lead-wire broken

Replace lead-wires

Lead-wires not connected 

properly

Reinsert lead-wires

correctly

Treatment time has expired

Switch unit OFF and  back ON

The battery power is low

Replace batteries

Unpleasant feeling beneath the pads

Poor pad contact

Ensure the pads have a thin layer 

of gel on them and they are flat 

against your skin

The unit does not turn on even though

new batteries have been inserted

Device may be faulty

Contact customer service

13

WL-2413B STIMULATOR TECHNICAL SPECIFICATIONS

Channel:

 Dual, isolated between channels.

Pulse amplitude:

 Adjustable 0 – 80mA peak into 500

Ω

 load each channel.

Pulse Rate:

 As pre-programming operation mode.

Pulse Width:

 As pre-programming operation mode.

Software ramp up feature:

 Pulse width ramp up when change mode.

Timer:

 30 minutes.

LCD:

 Show modes, pulse rate, pulse width, timer, CH1/CH2, intensity level, pause.

Function

Mode

Pulse Width 

(uS) 

Frequency 

(Hz) 

Ramp Up 

(sec)

On Time 

(sec)

Ramp Down 

(sec)

Off Time

(sec)

P1

300

50

2

5

2

9

P2

300

60

2

4

2

8

P3

300

70

2

3

2

7

Wave Form:

 Symmetrical Bi-Phasic square pulse.

Max charge per pulse:

 24 micro-coulombs maximum.

Power Source:

  3x1.5V(AAA) / 4.5 Volt batteries 

All electrical specifications are ±20% at 500

Ω

 load.

This symbol means “ Serial 
number “
      
  

These symbol mean 
“ Attention, consult 
the accompanying 
documents 

 

    

 

This symbol means
“ Manufacturer “ 
 

  

 

 

This symbol means type BF 
equipment; this device offers 
protection against electrical 

shock by stand compliance to leakage 
currents.
 
  

This symbol means please apply 
regulations enforced in your country   
when disposing your batteries and 
product at its end of life Ensuring it is  

disposed in an environmentally friendly 
manner.

Description of symbols : There are a number of technical symbols on your ems device explained as follows:

Summary of Contents for Contour

Page 1: ......

Page 2: ...Turning OFF the Device 8 Selecting the Program 9 Selecting the Exercise Intensity Level 9 SPECIAL FEATURES 10 CARE AND MAINTENANCE 11 TROUBLESHOOTING 13 WL 2413B STIMULATOR TECHNICAL SPECIFICATIONS 13 Warranty 14 Please read this Use and Care Manual carefully before you start using your TekSmart EMS Device Careful reading of the contraindications warnings precautions adverse reactions caution with...

Page 3: ...s has not been established You suffer from cancer The effects of electronic stimulation on cancerous tissue are unknown You are under medical supervision for cognitive dysfunction as you may not be able to comply with safety instructions The unit is in close proximity to shortwave or microwave diathermy equipment or you are connected to high frequency surgical equipment because of risk of device i...

Page 4: ...Do not use this system while sleeping Do not use if you feel numbness Do not use this system in or close to water Do not use the pads over or close to cancerous lesions Do not use the device in the following circumstances when there is a tendency to bleed internally following an injury of fracture following recent surgical procedures when stimulation may disrupt the healing process over areas of s...

Page 5: ...rience skin irritation or hypersensitivity due to the electrical stimulation or the conductive medium Keep the stimulator out of the reach of children Application of moderate heat thermal wrap to muscles as well as moistening skin prior to treatment improves treatment efficacy use of cold packs on treated muscles after treatment is likewise recommended This unit should only be used with the leads ...

Page 6: ...red as skin irritation and it will normally disappear within 30 minutes after the electrodes are removed If the redness does not disappear after 30 minutes from the removal of electrodes do not use the stimulator again until after the excessive redness has disappeared Turn off the stimulator if the stimulation feels unpleasant or does not provide pain relief Keep your system out of the reach of ch...

Page 7: ... your eyes and to your head and face You should stop using the device and should consult with your physician if you experience adverse reactions from the device Conditions that may affect your TekSmart EMS System Since the stimulator is a battery operated electronic system its output performance and safety may be affected greatly in extreme humidity Therefore it is very important to keep the stimu...

Page 8: ... 7 Time remaining 8 CH1 intensity level 9 CH2 intensity level 10 Lock status indicator 11 Battery status indicator 12 Pause indicator 13 Battery compartment 14 Battery cover 88 88 88 Pi8 PAUSE CH1 CH2 MIN MODE CH1 CH2 ON OFF 3 AAA Batteries EMS Device Lead Wires 4 Silicone Electrodes 3 4 oz Tube Conductive Gel Conductive Gel 3 4 oz The Curvassure Butt Toning Shorts ...

Page 9: ...electrode under the pocket lip 3 Insert pin end of the four lead wires through the eyelet hole on the outside of the shorts This can be done on the left side or right side depending on which pocket you prefer to put the device in during your session 4 Slide one Blue tip wire over and out of the upper right eyelet hole Slide the other Blue tip wire down and out of the lower right eyelet hole Slide ...

Page 10: ... In an emergency you may also pull the connector wire s from the device Note To prevent unpleasant electric shocks never remove the device while it is still turned on Selecting the Program The Device offers three different preset training programs The programs differ with respect to varying pulse widths frequencies and ON OFF time 1 Press the MODE button The P will be flashing Press ON or OFF repe...

Page 11: ...intensity level will reset to 0 each time the device turns off including after training sessions or changing the Program during training session Your Contraction Intensity Level For those not familiar with the sensation of ems it can be a little strange As you increase the intensity you will initially feel a tingle in the area under the pads Gradually increase the intensity until you feel your mus...

Page 12: ...ill flash on the display Press MODE again PAUSE will disappear on the display and session will continue The intensity level will stay at your current setting and time will continue to count down Automatic Shut off The device automatically turns off when no button is pressed for 60 seconds The device automatically turns off when the time for your exercise session has elapsed Intensity Level Reset F...

Page 13: ...ect the cables from the stimulator when not in use Do not pull on the cables but rather on the plugs attached to the ends of the cables Store the stimulator with the cables in a clean dry place Electrodes The electrode pads are conductive silicone pad reusable for single person use Take off from the garment after each use wipe off and dry the electrodes with a soft cloth dampened with soap and wat...

Page 14: ...gnetic immunity The device is intended for use in the electromagnetic environment specified below The customer or the user of the device should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood con crete or ...

Page 15: ...customer service 13 WL 2413B STIMULATOR TECHNICAL SPECIFICATIONS Channel Dual isolated between channels Pulse amplitude Adjustable 0 80mA peak into 500Ω load each channel Pulse Rate As pre programming operation mode Pulse Width As pre programming operation mode Software ramp up feature Pulse width ramp up when change mode Timer 30 minutes LCD Show modes pulse rate pulse width timer CH1 CH2 intensi...

Page 16: ... to be free from defects in workmanship and materials at the time of delivery The system accessories short and conductive gel for reasons of hygiene they can neither be taken back nor exchanged once used The distributor reserves the right to replace or repair the unit at their discretion HOW TO GET HELP To get assistance or answers to your questions please contact TekSmart Fitness LLC Visit our we...

Reviews: