22
4.1 Voltage Calibration
At the power supply displayed
2.000A
&
05.00V
in setting area of LCD and
**.**V
displayed in main
part of LCD. Connect the output terminal with an external reference voltage meter and shift to CV
mode. Hold the
LEFT
or
RIGHT
arrow keys to move the cursors to left or right and push the
UP
or
DOWN
arrow keys to adjust the values same as readings of external reference voltage meter, then
click
ENTER
button to finish the voltage bias calibration.
At the power supply displayed
2.000A
&
30.00V
in setting area of LCD, hold the
LEFT
or
RIGHT
arrow
keys to move the cursors to left or right and push the
UP
or
DOWN
arrow keys to adjust the values
same as readings of external reference voltage meter, then click
ENTER
button to finish the voltage
gain calibration.
4.2 Current Calibration
Push the
V/A
button and power supply displayed
0.500A
&
20.00V
in setting area of LCD and
*.***A
displayed in the main part of LCD. Connect the output terminals with the external reference current
meter and the load (6Ω/100W), shift to CC mode. Hold the
LEFT
or
RIGHT
arrow keys to move the
cursors to left or right and push the
UP
or
DOWN
arrow keys to adjust the values same as readings of
the external reference current meter, then click
ENTER
button to finish the current bias calibration.