21
IV Calibration
Follow the below chart, connect the 5 digit displayed volt meter, current meter, resistance
(10Ω/100W) into the output terminals, calibration follow the steps start from point of zero voltage -
voltage coefficient – current zero – current coefficient.
Hold the
SET
button to switch on power supply till “
REF
” displayed in LCD to enter calibration mode.
5
6