4) Press “

” button, you can use the relative function to eliminate the zero error. 

5) Connect the probes across capacitor to be tested. 

      Note: When testing 400

μ

F / 4000

μ

F capacitor, note that there will be approx. 30 sec time lag. 

Caution: Ensure that the circuit to be tested is "dead".   

Max. input over-load: 250V RMS. and<10sec. 

 

2.8 Frequency measurement 

1) Connect the black test lead to "

COM

" socket and red test leads to the "

 FREQ 

" socket. 

2) Set the selector switch to “

DUTY

 

FREQ

” position. 

3) Connect the probes across the source or load under measurement. 

 

   2.9 Temperature measurement 

1)

 

Set the selector switch to the ” 

TEMP °C/°F

” range and connect the k type thermocouple’s 

 

black    ( or “-“ terminal ) test lead to “

COM

” socket and red    ( or “+“ terminal ) test lead to the   

"

 TEMP 

" socket and read value on display screen. The range of TP-02 thermocouple is   

-50~+700°C ( -58 ~1292°F). 

2)

 

Press 

SELECT

 button to switch °C or °F display. 

3)

 

Opening the input will display the temperature at the meter terminals by the internal built-in   

sensor. 

     

   2.10 Auto Power Off  

When the meter has been turned on about 15 minutes without any action from the users, the 
meter will automatically change to “OFF” mode, and you can press any function button to 
wake up again.

 

If press and hold 

SELECT

 button to turn on the meter, the AUTO POWER 

OFF function will be canceled. 

 

   2.11 Using Communications  

 

1) Connecting the RS232 cable to the meter and a computer as shown in Figure 1, then turn on   

the meter. 

2) Press                  button more than 2 seconds on the panel to start the RS232 communication   

function. If you want to cancel the RS232 function, please press the                    button more than   
2 seconds once again. 

3) About the father information of communication, please refer to the on-line help of Data Logger   

attached. 

 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 

3.CARE AND MAINTENANCE 

Figure 1. RS232 Communication 

 

                                                                                               

 

6

Summary of Contents for DT9602R

Page 1: ...circuit Remember to place the meter in series with the circuit When servicing the meter use only specified replacement parts Use caution when working above 30 V ac rms 42 V peak or 60 V dc Such voltag...

Page 2: ...tage is below 7 5V approx Auto Power Off 15 minutes after stopping the switch or no any operation for the press button the meter automatically enters to power off mode Safety standards ETL EMC LVD The...

Page 3: ...nF 4 F 40 F 3 5 400 F 4000 F 20 5 Overload protection 250V DC AC RMS Frequency 9 999Hz 10MHz 0 1 rdg 5 Sensitivity 100kHz 1 5V RMS 100kHz 5V RMS Overload protect 250V RMS Duty Cycle 0 1 99 9 2 rdg 5 O...

Page 4: ...cycle capacitance diode continuity test and temperature measurement 3 Relative Measurement RS232 enable Press this button the meter enter relative measuring state and symbol appear on LCD The result...

Page 5: ...ce measurement 1 Connect the black test lead to COM socket and red test leads to the Ohm socket 2 Set the selector switch to Ohm CAP position and press SELECT button to select resistance measuring mod...

Page 6: ...e of TP 02 thermocouple is 50 700 C 58 1292 F 2 Press SELECT button to switch C or F display 3 Opening the input will display the temperature at the meter terminals by the internal built in sensor 2 1...

Page 7: ...ing new Do not use harsh chemicals cleaning solvents or strong detergents to clean the multimeter 6 Use only fresh batteries of the required size and type Always remove old or weak batteries They can...

Page 8: ...Screw Fuse 0 8A 250V FAST Fuse 10A 250V FAST Fuse Replacement 9602RCk507 2 8...

Reviews: