TekPower DT9602R Operation Manual Download Page 5

exit this mode and “ MAX ” / ” MIN ” symbol disappear on display. This button do not effect on 
diode and continuity test function, capacitance measurement, frequency measurement. 

 

     

2.2 Check the 9-volt battery.

   

      If the battery is weak, a “

 

 

" symbol will appear on the left of the display. It means that the 

battery should be replaced. 

 

   

  2.3 DC and AC Voltage measurement 

1)Connect the black test lead to "

COM

" socket and red test leads to the "

VOLT

" socket.   

2)Set the selector switch to “ 

VOLT DC 

” or “ 

VOLT AC

” position to select DCV or ACV measuring 

mode. 

3)Connect the probes across the source or load under measurement. 

 

   

  2.4 DC and AC Current measurement 

1)Connect the black test lead to "

COM

" socket and red test leads to the "

    µA 

mA

" socket.   

2)For measurement up to 600mA. Set the selector switch to “

AC/DC µA

 ” or “

AC/DC mA

 ” 

position and press 

SELECT

 button to select DC or AC current measuring mode. 

3)For current measurement from 600mA to 10A, connect the red test lead to "

10A

" socket. Set 

selector switch to “

AC/DC 10A

 ” position and press 

SELECT

 button to select DC or ACA 

measuring mode.   

4)Connect the probes across the source or load under measurement. 

Note:    Not to measure a current in a circuit with voltage more than 250V. 

 

   2.5 Resistance measurement  

1) Connect the black test lead to "

COM

" socket and red test leads to the "

 Ohm

 

" socket. 

 
2) Set the selector switch to " 

Ohm

 /    /    /        /       

CAP

 

" position, and press 

SELECT

 button to 

select resistance measuring mode. 

3) Connect the probes across circuit to be tested. 

Caution: Ensure that the circuit to be tested is "dead".   

Max. input over-load: 250V RMS. and<10sec. 

 
   

 

2.6 Diode, continuity test 

1) Connect the black test lead to "

COM

" socket and red test leads to the " 

  / 

 

" socket. 

2) Set the selector switch to    " 

Ohm / 

  / 

  /          CAP 

"  position and press 

SELECT

 button 

to select Diode 

 

or continuity 

 

test mode. 

3) Connect the black and red test probe to the cathode (-) and anode (+) ends of diode to be   

                tested respectively. 

4) Read the forward voltage drop (junction) value from the display. If reverse connected the   

       probes to diode, display shows over-load.

 

5) In Continuity test, the beeper sounds continuously if the resistance is less then  
120 

approximately.

 

 

Caution: Ensure that the circuit to be tested is "dead".   

Max. input over-load: 250V RMS. and<10sec. 

                                                           

2.7 Capacitance measurement   

1) Before testing, discharge the capacitor by shorting its leads together. Use caution in handing   

capacitors because they may have a charge on them of considerable power before 
discharging. 

2) Connect the black test lead to "

COM

" socket and red test leads to the "

 CAP 

" socket. 

3) Set the selector switch to      " 

Ohm / 

  / 

  /        CAP 

"  position and press 

SELECT

 button 

to select capacitance measuring mode. 

                                                                                               

 

5

Summary of Contents for DT9602R

Page 1: ...circuit Remember to place the meter in series with the circuit When servicing the meter use only specified replacement parts Use caution when working above 30 V ac rms 42 V peak or 60 V dc Such voltag...

Page 2: ...tage is below 7 5V approx Auto Power Off 15 minutes after stopping the switch or no any operation for the press button the meter automatically enters to power off mode Safety standards ETL EMC LVD The...

Page 3: ...nF 4 F 40 F 3 5 400 F 4000 F 20 5 Overload protection 250V DC AC RMS Frequency 9 999Hz 10MHz 0 1 rdg 5 Sensitivity 100kHz 1 5V RMS 100kHz 5V RMS Overload protect 250V RMS Duty Cycle 0 1 99 9 2 rdg 5 O...

Page 4: ...cycle capacitance diode continuity test and temperature measurement 3 Relative Measurement RS232 enable Press this button the meter enter relative measuring state and symbol appear on LCD The result...

Page 5: ...ce measurement 1 Connect the black test lead to COM socket and red test leads to the Ohm socket 2 Set the selector switch to Ohm CAP position and press SELECT button to select resistance measuring mod...

Page 6: ...e of TP 02 thermocouple is 50 700 C 58 1292 F 2 Press SELECT button to switch C or F display 3 Opening the input will display the temperature at the meter terminals by the internal built in sensor 2 1...

Page 7: ...ing new Do not use harsh chemicals cleaning solvents or strong detergents to clean the multimeter 6 Use only fresh batteries of the required size and type Always remove old or weak batteries They can...

Page 8: ...Screw Fuse 0 8A 250V FAST Fuse 10A 250V FAST Fuse Replacement 9602RCk507 2 8...

Reviews: