Maintenance
Coaxial RF Power Standards Instruction and Service Manual
5-8
I
O
POWER
TEGAM
Dual Type IV Power Meter Model 1806A
VOLTMETER
BRIDGE A
BIAS
VOLTAGE
ON
OFF
RVG
TEST
TEMP
READY
ERROR
VOLTMETER
BIAS
VOLTAGE
ON
OFF
RVG
TEST
TEMP
BRIDGE B
READY
ERROR
DVM or DMM
V+
V-
BIAS VOLTAGE
RF IN TEMP IN
SENSOR
TEGAM
Model F11XX
TEGAM
Model M11XX
DVM or DMM
V+
V-
Figure 5.6 Connection Diagram for Characterizing a Working Standard
3.
Connect the Reference (Terminating) RF Power Standard to the Sensor port of the Working
Standard (refer to Figure 5.6). To connect the standards, turn the feet on the Reference
Standard so they're all the way up. Support the mount with one hand while tightening the
nut on the male connector, taking care to get a good connection between the connectors.
Once a good connection has been made adjust the feet of the Reference Standard (while
still supporting with the other hand) so they will hold the weight. Make sure RF connectors
are clean and use proper torque wrench to tighten.
4.
Connect Signal Generator output to the input of the Working (Feedthrough) Power
Standard (refer to Figure 5.6). Make sure RF connectors are clean and use proper torque
wrench to tighten.
5.
Connect a voltmeter to the VOLTMETER connectors on the Type IV Power Meter that is
being used with the Reference Standard. Connect another voltmeter to the VOLTMETER
connectors on the Type IV Power Meter that is being used with the Working Standard
(refer to Figure 5.6).
6.
Make sure the output of the Signal Generator is off. Record the DC voltage from the Type
IV Power Meter that is being used with the Reference Standard as V
OFF1
and the DC
voltage from the Type IV Power Meter that is being used with the Working Standard as
V
OFF2
.