Operating Instructions
Coaxial RF Power Standards Instruction and Service Manual
3-2
General
The TEGAM RF Power Standards are designed to be employed as the fourth arm of a bridge
configuration. These units are designed as the sensing element in RF power calibration and
measurement systems. These systems employ other electronic elements to effect control of the
measurement routines. An example of this type of system is the TEGAM System II Power Sensor
Calibration System. Proper use of these standards and the TEGAM System II is further
documented in other TEGAM Operation and Service Manuals. The Working RF Power Standards
(Models F1125, F1130, and F1135) are feedthrough mount-splitter combinations are use to
calibrate thermoelectric, diode, and thermistor power sensors. The Reference RF Power Standards
(Models M1125, M1130, and M1135) are terminating thermistor mounts commonly used as a
reference standard for calibrating working standards. Both types of standards are used with a
precision DC Substitution Bridge such as found in the TEGAM Models 1806(A), 1805B, and 1830A.
Model F1125, F1130, and F1135 Connector Descriptions
Figure 3.1 Model F1125 Front View
RF IN connector
SMA female connector that connects to signal generator output. The RF power that comes in this
connector is applied equally to the DUT and the power standard. The fact that equal RF power is
applied to both the power standard and the DUT is what allows the determination the calibration
factor of the DUT. There is at least 6 dB of insertion loss in the RF IN path and as much as 10.5
depending on the specific model (see “Specifications” in Chapter I Instrument Description).
TEMP connector
4 pin mini-microphone connector that is used to connect the internal heater to a heater control
circuit found in TEGAM Models 1805B, 1806(A), 1830A, and 1820. RF Power Standards require at
least two hours of warm-up time to reach their operating temperature. A Temperature Control
cable is supplied with the Models F1125, F1130, and F1135.
BIAS VOLTAGE connectors
Binding posts used to connect the thermistor element to a DC Substitution bridge circuit as is
found in the TEGAM Models 1805B, 1806(A), 1830A, and 1804. The bridge operates on the
principal of DC substitution, so only DC voltages and currents are present at these terminals when
connected.