Instrument Description
Coaxial RF Power Standards Instruction and Service Manual
1-9
BIAS VOLTAGE TEMP
Model 1806
VOLTMETER
BIAS VOLTAGE TEMP
VOLTMETER
BRIDGE A
BRIDGE B
DVM or DMM
V+
V-
Power Meter
Power
Reference
Model M11XX
Figure 1.5 Measuring a 50 MHz Reference Output on a Power Meter using
a Terminating RF Power Standard (Model M1125, M1130, M1135)
BIAS VOLTAGE
TEMPERATURE
Model 1805B
AM Out (rear of 1805B)
AM Input
RF Output
Signal Generator
BIAS VOLTAGE
RF IN TEMP IN
SENSOR
Model F11XX
Model M11XX
BIAS VOLTAGE TEMP
Model 1806
VOLTMETER
BIAS VOLTAGE TEMP
VOLTMETER
BRIDGE A
BRIDGE B
DVM or DMM
V+
V-
Figure 1.6 Calibration of a Working Power Standard Using
a Reference Power Standard