Tecnologic THP 96 Operating Instructions Manual Download Page 8

The instrument must not to be switched off during the programming mode
otherwise the data inserted in the last session will not be memorized.

5.3 - PROGRAMMING OF INPUT LIMITS

If the instrument is provided with input for normalized signals 4..20 mA,
0..20 mA or 0..10 V it is necessary to set input limits to have the correct
measure indication. For example: if the probe to connect has a range of
0..100 bar, it will be necessary to program 0 on "Lci" parameter (start
scale) and 100 on "Hci" parameter (end scale). At same time, we have to
modify the same parameters to have the decimal point. Example:
instrument connected with relative humidity probe with 20..99 %RH as
range, to have the decimal point it will occour to program "dP" in "on"
mode, "Lci" at 20.0 and "Hci" at 99.9.

6 - DESCRIPTION OF PARAMETERS

Here following are described all the instrument parameters; pls. note that
some of them could  do not appear  or because are according to the kind of
 used instrument or because are automatically  not qualified, as they're not
necessary. 

d1

 - DIFFERENTIAL SWITCHING POINT1: Value between starting and

stopping of output OUT1.

d2

 - DIFFERENTIAL SWITCHING POINT2: Value between starting and

stopping of output OUT2.

db

 - NEUTRAL ZONE : Value which represents the semi-band of the

Neutral Zone, in the homonymous functioning mode, overlapping Set1.

LS1

 - MINIMUM SET1: Minimum possible Set point 1 value  or lower

limit of Set point 1.

LS2

 - MINIMUM SET2: Minimum possible Set point 2 value  or lower

limit of Set point 2.

HS1

 - MAXIMUM SET1: Maximum possible Set point1 value or higher

limit of Set point1.

HS2

 - MAXIMUM SET2: Maximum possible Set point2 value or higher

limit of Set point2.

Pb

 - PROPORTIONAL BAND : Amplitude of the band around Set 1 in

which proportional control takes place.

It

 - INTEGRAL TIME : Integral time to be set in algorithm of PID control

mode.

dt

 - DERIVATIVE TIME: Derivative time to set in algorithm of PID control

mode.

rSt

 - MANUAL RESET : Offset value for proportional band.

od

 - OUTPUTS DELAY: Delay from when the control need a variation to

the condition of the outputs to when this actually happens (expr. in sec.).
Non operating on output Out 1 in PID control mode.

Ct 

- CYCLE TIME: Cycle time of OUT 1 when proportional control takes

place in PID control mode.

drb

 - RESTART BAND  DYNAMIC SET: Semiband found in the symmetri-

cal zone in respect with Set outside of which intervenes the functioning of
the Dynamic Set Point.

dSI

 - DYNAMIC SET INCREMENT: Increment value of dynamic Set Point

function between two successive instant determines by the par. dSt.

dSt

 - DYNAMIC SET INCREMENT INTERVAL: Interval between one

Dynamic Set Point increment and the next (expr. in sec.).

Lci

 - LOWER NORMALIZED SIGNALS INPUT LIMIT: Value that the

instrument must to indicate in correspondance of minumum value (4 mA, 0
mA, 0 V). 

Hci

 - HIGHER NORMALIZED SIGNALS INPUT LIMIT: Value that the

instrument must to indicate in correspondance of maximum value (20 mA,
10 V).

CAL

 - CALIBRATION: Positive or negative offset which is calculated on

probe reading before visualizing and to which the control functioning is
also connected. This parameter can be utilized when a recalibration of the
instrument is desired.

Ft

 - CONTROL MODE: Allows the selection of one of possible control

modes that the instrument offers: ON/OFF (on), PID (Pi) or Neutral Zone
(nr).

PSE

 - INPUT PROBE: Allows selecting, if the instrument type is for

thermocouples or for thermoresistances, various types of probe: J (FE), K
(Cr), S (rh) for thermocouples and Pt100 (Pt), Ni100 (ni) for
thermoresistances. On changing this parameter it is to be remember that
the exit programming phase is to be carried out first afterwhich the
instrument is to be switched off and then on again.

OCO

 - SETS CONNECTION: Determines if the two Sets are to be

independent or if Set2 must be considered relative to Set1 (in=independent,
di=dependent).

HC1 

- OUTPUT 1 OPERATING MODE: Determines if the output OUT1 is

tocontrol a process which increment is intended as positive or reverse (ex.
Heating, Humidification etc.) or a process which increment is intended as
negative or direct (ex. Cooling, Dehumidification etc.) (H=reverse ,
C=direct).

HC2 

- OUTPUT 2 OPERATING MODE: Determines if the output OUT2 is

to control a process which increment is intended as positive or reverse (ex.
Heating, Humidification etc.) or a process which increment is intended as
negative or direct (ex. Cooling, Dehumidification etc.) (H=reverse ,
C=direct).

dP

 - DECIMAL POINT : Allows the insertion of the decimal point on the

display and therefore to determine resolution of the reading value (1 or
0,1) but not modified the Set, the Set limits (par. "LS", "HS") and the input
limits (par. "Lci", "Hci"). Example, if the Set was 20 the new Set will be 2.0
(on= with decimal point, oF=without decimal point).(available only for
models provided for PTC, RTD and normalized signals input). 

rou 

- UNIT OF MEASUREMENT: Determines the visualization of the

temperature in Centigrade or Fahrenheit degree. It is to be remember that
the change of this parameter modifies the visualization but not the Set and
the Set limit ("LS" and "HS") programmed (eg. if the Set was 50°C and the
unit changes, the Set will rest 50°F).

tun 

- AUTOTUNING: Determines the selection of Autotuning function for

the automatic setting of  PID parameters (n=no autotuning, y=start
autotuning).

hdd

 - HALF DIGIT DISPLAY: This parameter allows the approximation of

the last digit at right side. Infact when this parameter is in on mode, on this
digit will be show 0 when the real measuring value is between 0 and 4, and
will be show 5 when the real measuring value is between 5 and 9. For
example if the real measure is 78 display will show 75, or if the real
measure is 70.3 the display will show 70.0 (n=without approximation,
y=with approximation).

tAb

 - FIXED PARAMETER

6.1 - PARAMETERS

 

TABLE

0

-999 ... +999

Calibration

CAL

999

-999 ... 7000

Higher limit for normalized
signal input

Hci

-99

-999 ... 7000

Lower limit  for normalized
signal input

Lci

1

1 ... 3600 sec.

Dynamic Set increment
interval

dSt

0

0 ... Probe limit

Dynamic Set increment

dSI

0

0 ... Probe limit

Restart band Dynamic Set

drb

20

1 ... 500 sec.

Cycle time

Ct

0

0 ... 500 sec.

Outputs delay

od

0

Probe limit

Manual reset

rSt

30

0 ...3600 sec.

Derivative time

dt

300

0 ...3600 sec.

Integral time

It

40

1/0,1 ... Probe

limit

Proportional band

Pb

Max. 

Probe limit

High/Maximum set 2

HS2

Max. 

Probe limit

High/Maximum set 1

HS1

min. 

Probe limit

Low/minimum set 2

LS2

min. 

Probe limit

Low/minimum set 1

LS1

1

1/0,1 ... Probe

limit

Neutral zone

db

-1

Probe limit

Differential 2

d2

-1

Probe limit

Differential 1

d1

Notes

Def.

Range

Description

Par.

TECNOLOGIC - THP 96

 

 USER MANUAL (I - GB) - Vr. 02   - ISTR 00997  - PAG. 8

Summary of Contents for THP 96

Page 1: ... exclusive property of TECNOLOGIC S p A which forbids any reproduction and divulgation although partial if not expressly authorised TECNOLOGIC S p A reserves the right to execute aesthetically and functional modifications at any moment and without any notice INDICE PROBLEMI MANUTENZIONE E GARANZIA 7 TABELLA PARAMETRI 6 1 DESCRIZIONE DEI PARAMETRI 6 IMPOSTAZIONE DEI LIMITI DI INGRESSO 5 3 PROGRAMMA...

Page 2: ...5 C Umidità ambiente di funzionamento 30 95 RH senza condensazione Temperatura di trasporto e immagazzinaggio 10 60 C CARATTERISTICHE FUNZIONALI Regolazione ON OFF ZONA NEUTRA PID Range di misura Secondo la sonda utilizzata vedi tabella Risoluzione visualizzazione Secondo la sonda utilizzata 1 0 1 5 0 5 Precisione totale 0 5 fs Velocità di acquisizione 1 acquisizione al secondo Azione tipo 1C seco...

Page 3: ...di ciclo dell uscita Per il termine INTEGRALE It Tempo Integrale Per il termine DERIVATIVO dt Tempo derivativo Lo strumento è impostato in fabbrica con i parametri relativi alla regolazione PID su valori standard Qualora questi non dovessero risultare ottimali per la regolazione si consiglia di attivare la funzione di AUTOTUNING che permette la sintonizzazione dello strumento per la regolazione PI...

Page 4: ...INIMO 2 Valore minimo impostabile come Set2 o limite inferiore del Set2 HS1 SET MASSIMO 1 Valore massimo impostabile come Set1 o limite superiore del Set1 HS2 SET MASSIMO 2 Valore massimo impostabile come Set2 o limite superiore del Set2 Pb BANDA PROPORZIONALE Ampiezza della banda intorno al Set1 nella quale interviene la regolazione proporzionale It TEMPO INTEGRALE Tempo integrale da impostare ne...

Page 5: ...messaggi di errore sonda EEEE in caso la sonda sia interrotta o in overrange oppure nel caso la sonda PTC RTD o per segnali normalizzati sia in cortocircuito o in underrange in tal caso verificare la corretta connessione della sonda con lo strumento e successivamente procedere alla verifica della stessa Nelle condizioni di errore tutte le uscite sono disattivate PULIZIA Si raccomanda di evitare l ...

Page 6: ...e per second Action 1C type according to EN 60730 1 Compliance ECC directive EMC 89 336 EN 50081 1 EN 50082 1 ECC directive LV 73 23 and 93 68 EN 60730 1 MEASUREMENT RANGE TABLE 99 9 700 0 999 7000 4 20 mA 0 20 mA 0 10 V gener 0 1600 C 32 2912 F tc S rh 0 1200 C 32 2192 F tc K Cr 0 800 C 32 1472 F tc J FE 50 0 150 0 C 58 0 302 0 F 50 150 C 58 302 F Ni 100 ni 99 9 600 0 C 99 9 999 9 F 100 600 C 148...

Page 7: ...of 4 hours after this time if the Autotuning is not completed the instrument automatically get out from the procedure showing constantly on the display the EEE indication To stop the Autotuning cycle or re establish the normal functioning after an error switch off and on the instrument The calculated values will be automatically memorised by the instrument at the end of the Autotuning cycle in the...

Page 8: ... selection of one of possible control modes that the instrument offers ON OFF on PID Pi or Neutral Zone nr PSE INPUT PROBE Allows selecting if the instrument type is for thermocouples or for thermoresistances various types of probe J FE K Cr S rh for thermocouples and Pt100 Pt Ni100 ni for thermoresistances On changing this parameter it is to be remember that the exit programming phase is to be ca...

Page 9: ...ivated HOW TO CLEAN We recommend to avoid abrasive cleaners or containing solvents which could damage the instrument WARRANTY AND REPAIRS The instrument is under warranty against construction vices or defected material noticed within 12 months from delivery date The warranty is limited to the repairs or to the substitution of the instrument The eventual opening of the housing the violation of the ...

Reviews: