background image

IST-2237.EC01.01

 

Istruzione / User’s Manual / Manuel d’utilisation 

Pag.20/20

 

 

 

 

TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) 

Tel. +39 02 26922890 - Fax +39 02 2133734 

 

Tabella 13 / Table 13 / Tableau 13  SE237ECL (

Cloro / Chlorine / Chlore

“S1-SET” 

(Dip-Switch)

 

Livelli Allarme (

Relé

Alarm levels(

Relays

) / Niveaux d'alarmes(

Relais

)

 

Ritardo dei Relé (

in Secondi

Relays Delay (

Seconds

) / retarde du relais (

Secondes

)

 

 

ALARM 1 

ALARM 2 

ALARM 3 

ALARM 1 

ALARM 2 

ALARM 3 

FAULT

 

OFF 

OFF 

OFF 

0,3 ppm 

0,4 ppm 

0,5 ppm 

12 

30 

60 

30 

ON 

OFF 

OFF 

0,5 ppm 

0,7 ppm 

1,0 ppm 

12 

30 

60 

30 

OFF 

ON 

OFF 

0,5 ppm 

0,8 ppm 

1,2 ppm 

12 

30 

60 

30 

ON 

ON 

OFF 

0,5 ppm 

1,0 ppm 

1,0 ppm 

12 

30 

60 

30 

OFF 

OFF 

ON

 

TLV-TWA

 

(COSHH)

 

0.5 ppm 

TLV-STEL

 

(COSHH)

 

1.0 ppm 

TLV- C

 

1.0 ppm

 

30 

ON

  

OFF 

ON

 

Risevato 

/ Reserved / Réservé

 

Risevato 

/ Reserved / Réservé

 

OFF 

ON

  

ON

 

Risevato 

/ Reserved / Réservé

 

Risevato 

/ Reserved / Réservé

 

ON 

ON 

ON 

Risevato 

/ Reserved / Réservé

 

Risevato 

/ Reserved / Réservé

 

COSHH

 

= Ente Europeo / European Department / Organisme européen 

OSHA =

 

Ente Americano / U.S. Department / Organisme Etats-unien 

 

Fig. 1 –

 

Dimensioni /

 Size / 

Dimensions 

 

 

1

 

2

 

3

 

4

 

ON

 

OFF

 

Centrale rivelazione gas 

Gas Central Unit

 

Centrale de détection

 

RL

 

-

 

+

 

Dip Switch

 

mV

 

Test Uscita  

Output Test 

Sortie Test 

40÷200mV

 

 

 

 

 

Uscita in 4÷20 mA 

4÷20 mA Output 

Sortie en 4÷20 mA

 

 

 

ALARM 1 

 

 

ALARM 2 

 

 

 

 

ALARM 3 

 

 

FAULT 

SET

 

Fig. 2 -

 Schema di collegamento

 

/ Wiring diagram / Schéma

 

TESTmV

 

TP2 

TP1 

Scheda Uscite Relé / Relays Output Board / Sortie sur Realis

 

12

-2

4V

 

m

A

 

12÷24Vdc 

Alimentazione 

Power Supply 

Alimentation

 

 

 

Fig. 3 -

 Tester di calibrazione 

/ Calibration Tester / Kit de Calibration 

F1 

F2 

Tasti / 

Key / Touches

 

TC014

 

0,3 L/min

 

1 L/min

 

 

 

Alimentazione 

/ Power Supply / Alimentation

 

100

 

300

 

400

 

500

 

50

 

0

 

5

 

200

 

Vdc

 

RL=Resistenza di carico 

/ Load resistor / Resistance de charge

 

RL

 

Ohm

 

Fig.4 -

 

Alimentazione / Resistenza di Carico 4÷20mA 

Power supply / Load resistance diagram 4÷20mA 
Alimentation / Résistance de charge 4÷20mA

 

600

 

0

 

10

 

15

 

20

 

25

 

28

 

650

 

10,8

 

24

 

Area di funzionamento 

Allowable oparating region 

350

 

700

 

800 

 

Summary of Contents for SE237E Series

Page 1: ...re Caratteristiche tecniche Technical specifications Caractéristiques techniques 2 IT DESCRIZIONE 4 NOTE SUL GAS RILEVATO DAI VARI MODELLI 4 FUNZIONAMENTO 5 INSTALLAZIONE 6 AVVERTENZE 6 VERIFICHE E CALIBRAZIONE 6 EN DESCRIPTION 8 NOTES ON THE DETECTED GAS 8 OPERATIONAL DESCRIPTION 9 INSTALLATION 10 WARNING 10 TEST and CALIBRATION 11 FR DESCRIPTION 12 MODÈLES ET CARACTÉRISTIQUES DU GAZ DÈTECTEÈ 12 ...

Page 2: ...chelle 500 ppm 500 ppm 1000 ppm 1000 ppm 150 ppm 500 ppm Vita media in aria pulita Average Life in fresh air Vie moyenne en air pur 2 anni years ans 2 anni years ans 3 anni years ans 2 anni years ans 2 anni years ans 2 anni years ans Tempo massimo di immagazzinamento Max Storage Time Temps maximum de stockage 3 mesi month mois 1 mese month mois 3 mesi month mois 3 mesi month mois 3 mesi month mois...

Page 3: ...th mois 2 mese month mois 3 mese month mois 2 mese month mois Risoluzione Resolution Résolution 0 1 ppm 1 ppm 1 ppm 0 7 ppm 0 1 ppm Tempo di risposta T90 Response Time T90 Temps de réponse T90 120 secondi Seconds secondes 60 secondi Seconds secondes 60 secondi Seconds secondes 2 70 secondi Seconds secondes 50 secondi Seconds secondes Ripetibilità Repeatability Répétitivité 2 del segnale of the sig...

Page 4: ... persone dai polmoni raggiunge il sangue dove impedisce il trasporto dell ossigeno ai tessuti e agli organi SE237EC H è identico al precedente ma utilizza un sensore di livello superiore adatto a industrie e ambienti con presenza d inquinati e altri gas tossici SE237ES ha un sensore in grado di rilevare Biossido di zolfo SO2 sinonimi Anidride solforosa Ossido di zolfo Ossido solforoso è un gas tos...

Page 5: ... ALARM 2 interviene dopo 30 secondi oppure immediatamente quando è superato il livello TLV STEL Il 3 Led rosso ALARM 3 si accende se la concentrazione di gas supera il 3 livello d allarme impostato Il relè ALARM 3 interviene dopo 60 secondi oppure immediatamente quando è superato il livello TLV C I TLV threshold limit values sono i valori limite d esposizione a sostanze inquinanti a cui i lavorato...

Page 6: ...di Funzionamento Guasti Importante terminata l installazione alimentare l apparecchio attendere circa 20 30 minuti e poi se necessario per adattare il sensore alle condizioni ambientali eseguire la Regolazione dello Zero vedi Verifiche e Calibrazione AVVERTENZE La vita utile del sensore in aria pulita è mediamente di 2 anni Vedi a pag 1 Al termine di questo periodo indi cato dallo strumento con un...

Page 7: ...5ppm NH3 4 8 mA 15ppm CO 6 4 mA 3ppm SO2 4 8 mA 5ppm H2S 8 mA 2 5ppm HCN 5 6 mA 10ppm NO 6 6 mA 5ppm NO2 8 mA 2 5ppm HCL 4 8 mA 0 5ppm HCL CALIBRAZIONE Codice F2 F2 F2 F1 F2 F1 serve per ritarare il sensore con la miscela di gas sopra indicata AVVISO per garantire la sicurezza che non avvengano errori d elaborazione esiste la rara possibilità che durante la Calibrazione il Led Giallo si spenga ogn...

Page 8: ...and mucous membranes respiratory and digestive tracts Ammonia is used in industrial refrigeration applications It is also used as fertilizers either as its salts as a component in paints for the production of plastics and poly mers such as solvent or as household cleaners SE237EA H model is as previous but it uses a sensor suitable for industry and polluted environments and with presence of other ...

Page 9: ...sensor needs a time of preliminary heating of about 60 seconds During this peri od the yellow LED FAULT flashes After this period the yellow LED light off the green LED ON illuminates to in dicate normal functioning After this period the unit is able to detect gas even if it attains the optimum stability con ditions after about 4 hours continual functioning Normal operation the green LED ON should...

Page 10: ...ls on the main board are plug in type it is necessary to extract them to make the connec tion pay attention when you insert them again being polarized The relays terminals on outputs board are fixed The sealed relays are tension free SPST Single Pole Single Throw contacts and should be set NO Normally Open or NC Normally Closed by positioning the 4th Dip Switch see Table 2 With the Dip Switches fr...

Page 11: ...ration every 12 months according to the use Note this function is not working if the 1 and or the 2 and or the 3 red Led have already turned on ZERO ADJUST Zero Code F2 F1 F1 F2 this function is to adjust the Zero sensor and can be done in clean air only environment without the presence of toxic gas or other pollutants Consider that the SE237E series have a sophisti cated follower of zero that eve...

Page 12: ...ecteurs ponctuel pour gaz toxiques équipée d un capteur a cellule électrochimiques Le dé tecteurs peut être utilisé en détecteur ponctuel avec asservissements incorporés ainsi qu en systèmes centralisés d alarme pour parkings laboratoires industries et environnements qui doit être protégé de la présence anormale de gaz Le détecteur est constitué par un boîtier contenant le circuit électronique et ...

Page 13: ...ustion en partie par les émissions naturelles des éruptions volcaniques des incendies ou des activités civiles et industriels de combustion tels que les transports véhicules à moteur diesel essence GPL etc et la production du chaleur et d électricité Le SE237EN2 permet de détecter le Dioxyde d azote NO2 est un gaz toxique d odeur âcre brun rougeâtre plus lourd que l air Sa densité relative à l air...

Page 14: ...e la remplacer par une nouvelle 3 si une nouvelle Car touche est installée ou bien elle n est pas correctement connectée ou bien elle n est pas compatible Contrôler les connexions avec la Cartouche et la compatibilité Page 1 Exécutez le contrôle en coupant et en rétablissant l alimentation du détecteur Si la condition perdure il sera nécessaire de remplacer et ou de renvoyer le détecteur au fourni...

Page 15: ...ésents en quantités élevées ils peuvent interférer avec le capteur il est conseillé de ventiler l ambiance après emploi de ces substances Considérer qu en ambiances particulièrement polluées par des vapeurs de solvants organiques hydro gène alcool en quantités élevées des fausses alarmes peuvent apparaître et que le capteur devra pour retrouver ses conditions de fonctionnement normales rester long...

Page 16: ...voyer le détecteur au fournisseur pour réparation Important Durant l opération d étalonnage l appareil porte la sortie S à 0mA L étalonnage doit être exécuté en air propre exclusivement milieu sans la présence de gaz toxiques ou autres A l aide des touches exécuter le Code d Etalonnage Attendre que les LED jaune et verte s allument en feu fixe et que la 1ère LED rouge clignote Coiffer la tête de d...

Page 17: ...installation du détecteur sous 12Vcc Massima distanza cui può essere installato ogni rivelatore se è alimentato a 24Vcc The maximum distance to install each detector from the 24Vdc power Supply Distance maxi d installation du détecteur sous 24Vcc 0 75 mm2 26 Ω km 100 m 300 m 1 mm2 20 Ω km 150 m 400 m 1 5 mm2 14 Ω km 200 m 500 m 2 5 mm2 8 Ω km 400 m 800 m Tabella 2 Table 2 Tableau 2 S1 SET Dip Swit...

Page 18: ... 12 30 60 30 ON ON OFF 50 ppm 100 ppm 200 ppm 12 30 60 30 OFF OFF ON TLV TWA COSHH 25 ppm TLV STEL COSHH 35 ppm TLV C 50 ppm 0 0 0 30 ON OFF ON Risevato Reserved Réservé Risevato Reserved Réservé OFF ON ON Risevato Reserved Réservé Risevato Reserved Réservé ON ON ON Risevato Reserved Réservé Risevato Reserved Réservé Tabella 6 Table 6 Tableau 6 SE237EC S SE237EC H Monossido di carbonio Carbon mono...

Page 19: ...ays Delay Seconds retarde du relais Secondes 1 2 3 ALARM 1 ALARM 2 ALARM 3 ALARM 1 ALARM 2 ALARM 3 FAULT OFF OFF OFF 25 ppm 50 ppm 100 ppm 12 30 60 30 ON OFF OFF 25 ppm 50 ppm 75 ppm 12 30 60 30 OFF ON OFF 15 ppm 25 ppm 50 ppm 12 30 60 30 ON ON OFF 15 ppm 35 ppm 60 ppm 12 30 60 30 OFF OFF ON TLV TWA OSHA 25 ppm TLV STEL OSHA 25 ppm TLV C OSHA 50 ppm 0 0 0 30 ON OFF ON Risevato Reserved Réservé Ris...

Page 20: ...servé Risevato Reserved Réservé COSHH Ente Europeo European Department Organisme européen OSHA Ente Americano U S Department Organisme Etats unien Fig 1 Dimensioni Size Dimensions 1 2 3 4 ON OFF Centrale rivelazione gas Gas Central Unit Centrale de détection RL Dip Switch mV Test Uscita Output Test Sortie Test 40 200mV Uscita in 4 20 mA 4 20 mA Output Sortie en 4 20 mA ALARM 1 ALARM 2 ALARM 3 FAUL...

Reviews: