![Tecno-gaz TD010TMD Service Manual Download Page 19](http://html1.mh-extra.com/html/tecno-gaz/td010tmd/td010tmd_service-manual_1071770019.webp)
Page
19
of
52
Ref.
Parameter
Name
Min
Max
Default
Menu
3131
Menu
3211
#25.01
Cold
water
loading
0=
Off
1=
On
NA
#25.02
Hot
water
loading
0=
Off
1=
On
NA
#25.03
Demineralised
water
loading
0=
Off
1=
On
NA
#25.04
Prewashing
time
0
999
NA
Sec
PROGRAM
MENU:
WASHING
PHASE
Ref.
Parameter
Name
Min
Max
Default
Menu
3131
Menu
3211
#26.01
Cold
water
loading
0=
Off
1=
On
NA
#26.02
Hot
water
loading
0=
Off
1=
On
NA
#26.03
Demineralised
water
loading
0=
Off
1=
On
NA
#26.04
Detergent
enabling
0=
Off
1=
On
NA
#26.05
Neutraliser
enabling
0=
Off
1=
On
NA
#26.06
Rinse
aid
enabling
0=
Off
1=
On
NA
#26.07
Water
heating
element
enabling
0=
Off
1=
On
NA
#26.08
Detergent
enabling
temperature
0
99
NA
°C
#26.09
Neutraliser
enabling
temperature
0
99
NA
°C
#26.10
Rinse
aid
enabling
temperature
0
99
NA
°C
#26.11
Washing
time
0
999
NA
Sec
#26.12
Heating
element
activation
delay
time
0
999
NA
Sec
#26.13
Target
temperature
0
99
NA
°C
PROGRAM
MENU:
RINSING
PHASE
Ref.
Parameter
Name
Min
Max
Default
Menu
3131
Menu
3211
#27.01
Cold
water
loading
0=
Off
1=
On
NA
#27.02
Hot
water
loading
0=
Off
1=
On
NA
#27.03
Demineralised
water
loading
0=
Off
1=
On
NA
#27.04
Detergent
enabling
0=
Off
1=
On
NA
#27.05
Neutraliser
enabling
0=
Off
1=
On
NA
#27.06
Rinse
aid
enabling
0=
Off
1=
On
NA
#27.07
Water
heating
element
enabling
0=
Off
1=
On
NA
#27.08
Conductivity
probe
enabling
0=
Off
1=
On
NA
#27.09
Detergent
enabling
temperature
0
99
NA
°C
#27.10
Neutraliser
enabling
temperature
0
99
NA
°C
#27.11
Rinse
aid
enabling
temperature
0
99
NA
°C
#27.12
Rinsing
time
with
neutraliser
0
999
NA
Sec
#27.13
Heating
element
activation
delay
time
0
999
NA
Sec
#27.14
Target
temperature
0
99
NA
°C
#27.15
Conductivity
probe
drain
pump
ON
time
0
999
NA
Sec
#27.16
Conductivity
probe
drain
pump
OFF
time
0
999
NA
Sec
#27.17
Rinsing
time
without
neutraliser
0
255
NA
Sec
PROGRAM
MENU:
DISINFECTION
PHASE
PROGRAM
MENU:
DISINFECTION
PHASE
Ref.
Parameter
Name
Min
Max
Default
Menu
3131
Menu
3211
#28.01
Cold
water
loading
0=
Off
1=
On
NA
#28.02
Hot
water
loading
0=
Off
1=
On
NA
Summary of Contents for TD010TMD
Page 47: ...Page 47 of 52 DIAGRAMS...
Page 48: ...Page 48 of 52...
Page 49: ...Page 49 of 52 TD010TMD TD012TMD...
Page 50: ...Page 50 of 52 TD011TMD...