DDIFF051_REV1.doc
19 - 29
PAR 07
ANALYSIS AND RESEARCH OF FAILURES
N° FAULT
CAUSE
INTERVENTION
FEEDING CABLE NOT CONNECTED TO
THE SOCKET OR AT THE BASIS OF
THE MACHINE
CORRECTLY CONNECT
THE FEEDING CABLE
1
DOESN’T PRODUCE WATER
RIARMO NON ESEGUITO
PREMERE ILTASTO DI
RIARMO
YOU HAVEN’T PUT IN WATER RIGHT
TO THE MAX LECVEL
FILL THE DISTILLER
CORRECTLY
THE GASKET OF THE COVER IS NOT
CORRECTLY PLACED OR DOESN’T
SEAL ENOUGH
PLACE THE GASKET
CORRECTLY
3
DOESN’T PRODUCE ENOUGH
DISTILLED WATER DURING
THE CYCLE
THE WATER CONTAINER IS NOT
PLACED UNDER THE DRAINING OF
THE DISTILLER
PLACE THE CONTAINER
CORRECTLY
ACTIVATED COAL FILTERS ARE NOT
PLACED IN THE SPOUT
REPLACE FILTERS
CORRECTLY
4
LEACK OF WATER AND
STEAM FROM THE DRAIN
THE CONTAINER IS NOT PERFECTLY
ALIGNED WITH THE DISTILLER
ALIGNE THE CONTAINER
Summary of Contents for 250-S