DDIFF051_REV1.doc
18 - 29
PAR 06
ORDINARY AND EXTRAORDINARY MAINTENANCE
•
As for the external cleaning of the DISTILLER, you can use a neutral detergent and then dry.
•
Clean, if necessary, the gasket of the upper cover, but don’t remove it, it’s placed in a specific
direction (with the lip upwards)
•
The stainless steel internal chamber is delicate, don’t use abrasive detergents for cleaning
CLEAN THE MACHINE AT THE END OF EVERY CYCLE
1. Put hot water in the chamber just to cover the residues
2. Put from 1 to 4 teaspoons of detergent supplied inside the chamber, the quantity depends on the
residues.
3. With the chamber opened connect the feeding, bring to boiling the water and disconnect the
connection to the network
4. After 20 minutes drain the water and rinse with clean water
5. Repeat this operation until the chamber is completely clean
6.
Little quantities of residue in the bottom don’t have to be considered harmful, the machine works in
effective way.
FIG 4
COOLING COIL
If the cooling coil is obstructed, the effectiveness
of the distiller could be compromised.
For the cleaning, unscrew the screws (G) in the
part below the cover.
Remove the cover being careful during the
extraction to the joint in position (H).
Remove the coil and brush the cooling
commutators with care (I)
•
Don’t plunge the engine in the water
during the cleaning of the cover.
•
Add few drops of oil on the output shaft
in homogenous way.
FIG 5
Summary of Contents for 250-S
Page 2: ......
Page 8: ......
Page 16: ...DDIFF051_REV1 doc 13 29 PAR 09 SCHEMISTICA ESPLOSO file XDIM001...
Page 24: ...DDIFF051_REV1 doc 21 29 PAR 09 DIAGRAMS EXPLODED VIEW file XDIM001...
Page 32: ...DDIFF051_REV1 doc 29 29 PAR 09 SCH MA file XDIM001...
Page 33: ...Note...
Page 34: ...Note...
Page 35: ......