IST-2293.PX03.01
Istruzione / User’s Manual / Manuel d’utilisation
Pag.6/12
TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI )
Tel: 02/26 92 28 90 Fax: 02/21 33 734
“CALIBRATION” (
Calibration Code
: F2, F2, F2, F1, F2, F2)
:
this function allows completely recalibrate the sensor.
Important:
please, only
using a mixture 20%LEL (0.88%v/v)
Methane in Air (
20.9% Oxygen
)
Warning
: to guarantee that no errors of elaboration happen, the rare possibility exists that during the Calibration the
yellow LED switch off every 8 seconds,
in this case interrupt the procedure, switch off and witch on the instrument and
repeat the Calibration. If condition persists it will be necessary to send the detector to the supplier for the reparation.
Important note:
During Calibration routine the mA output indicates 0mA
.
The Calibration can be done in clean air only (
environment without the presence of flammable or other polluting gas
). With the
keys perform the “Calibration
Code
". Wait until the Yellow and Green Led switch on fix and the red Led starts to
flash. Insert the TC011 into the sensor holder to regulate the influx of the gas the way that the flow meter indicates
around 0.3 l/mins (
see Fig.3
). Wait around 3 minutes, then when the red Led switch on (
and while it's SWITCHED ON
),
press the key F2 on the instrument and hold it pressed until the Red Led is switched off for at least 2 seconds (
if the
Red Led continue to flash, repeat the operation
).
Then, close the gas cylinder and remove TC011
. At this point we can have
two possibilities:
Yellow and Green Led illuminates:
the calibration routine has correctly been performed.
Wait 8 seconds, until
the instrument automatically restores the normal working conditions
.
(
see “Operational Description > Preheating”
)
Yellow LED illuminates:
the routine has failed. In this case, wait 8 seconds, until the instrument automatically re-
peat
Preheating
, then repeat the “
Calibration”
routine without inserting again the code. If condition still persists after the
replacement of the cartridge, it will be necessary to send the detector back to the manufacturer for reparation.
“CALIBRATION CHECK” (
no Code required
)
:
this operation allow effecting a real functional test of the equipment
with gas
after the "
Calibration
" routine, or after the installation. The “
Calibration Check
” routine should be done dur-
ing the periodic maintenances because this is the only method to verify the instrument real functioning.
This routine should be done
using a mixture 20%LEL (0.88%v/v)
Methane in Air (
20.9% Oxygen
)
Insert the TC011 over the sensor holder, adjust the sample gas bottle valve as the flow meter indicates around 0.3
l/mins (
see Fig.3
),
verify with a voltmeter connected to the Test-Point “
TESTmA
”,
(
see fig.2
) the value reach the mV
value as shown in
Table 3, 4 and 5 on column 6
[corresponding to mA output value (
±0.2mA
) as indicated in
column 7
.
T
he central unit should be display about the %LEL value (±2) as indicated in
column 8
].
If the result is different, is
necessary to recalibrate the sensor (
see “Calibration”
).
Then, close the gas bottle, remove TC011.
Then, the mA out-
put will slowly decrease up to 4mA.
Using a cylinder with the specific gas/Air mixture, the output signal will correspond to the gas used. (Example if
you have to verify a hydrogen detector, using a cylinder with H
2
at 0.8% v/v (20% LIE) in air, the mA output will be
approximately 7.2 mA (corresponding about to 72mV measured on "
TESTmA
”).
F
DESCRIPTION
Le TS293P est un transmetteur (4÷20mA) pour gaz et vapeurs combustibles avec capteur catalytique Pellistor
pouvant également être utilisé en systèmes centralisés d'alarme pour l’industrie et le tertiaire. Le détecteur est
constitué par une un boîtier antidéflagrant contenant le circuit électronique et les borniers de raccordement; dans
le porte capteur, placé dans la partie inférieure du boîtier, est logée une "
cartouche capteur échangeable
" conte-
nant l'élément sensible et les données identificatrices et de réglage.
Le TS293P est un transmetteur sur 3 fils avec un signal de sortie
S
4÷20 mA avec fond d’échelle à 100% de la
LIE (
Limite Inférieure d'Explosivité
), du gaz mesuré (
Tableau 3
) Il s’utilise en se raccordant sur les centrales d’alarmes
mono et multivoies Tecnocontrol. Sur le couvercle, se trouvent les touches de codage F1 et F2 pour les opérations
de vérification et calibrage et 3 LED indiquant les conditions de fonctionnement:
Signalisations et commandes:
Led Rouge "ALARM":
signal d'alarme ALARM (20%LIE).
Led Vert "ON":
fonctionnement normal.
Led Jaune "FAULT":
capteur en panne ou déconnecté ou saturé ou échu.
Touches F1
et
F2
pour les opérations de vérification et calibrage
MODÈLES
Les données LIE des Gaz sous indiqués sont rapportés à la norme EN 61779-1.
Le
TS293PE
permet de détecter le
Acetilene (C
2
H
2
).
C'est un gaz incolore, plus léger que l'air, très infiammable.
Avec l'air former des mélanges explosifs. Sa densité relative à l'air est 0,9 et sa LIE est 2,3%v/v
exprimé en % Volume.
Le
TS293PX-S
à la demande, est étalonnable pour les gaz indiqués au
Tableau 3
et est utilisé dans des environne-
ments pollués faible. Si dans le milieu sont présents des gaz non indiqués dans le tableau on conseille l’utilisation
de l’appareil TS293PX-H.
Le
TS293PS
permet de détecter le
Styrène (C
8
H
8
).
Il est un liquide huileux jaune transparent, qui s'évapore à la
température ambiante, les vapeurs sont plus lourdes que l'air, toxique et inflammable. Sa densité relative à l'air est
3,6 et sa LIE, est 1,1% v/v, exprimé en % Volume. Son utilisation principale est comme monomère pour la produc-
tion de nombreuses matières plastiques.
Le
TS293PX-H
à la demande, est étalonnable pour les gaz indiqués au
Tableau 4
et est utilisé dans des environne-
ments pollués.