background image

IST-2183.KM01.02/B 

Istruzione / User’s Manual / Manuel d’utilisation 

Pag. 6 / 11 

 

 

 

TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI ) 

 Tel: +39 02 26 92 28 90   Fax: +39 02 21 33 734

 

 

Compatibility  with  other  Control  Units: 

In case  of a  control  unit  other  than  Tecnocontrol,

 

please  use  products 

with 4÷20mA input referred to ground and with compatible operating characteristics (Full Scale, Minimum operating 
voltage, Absorption, Maximum load resistance, etc.)

We accept no liability for any malfunction, failure or dam-

age caused by products that are not compatible or not of our production. 
Average life

: In fresh air and in normal working condition the sensor's life is 

about 5 years

 from the date of instal-

lation. After this period the yellow LED “

FAULT

” flashes every 4 seconds, is necessary replacing the detector. 

After  an  exposure  to  gas,  above  the  limits  of  measurement  (50%  LFL),  the  operation  of  the  detector  must  be 
checked with the titled gas, as explained in paragraph "

CALIBRATION CHECK

". 

Important:

 

The catalytic  sensor operates only in  presence of Oxygen.  Do  not use pure gases or  a lighter 

directly on the sensor since they could damage it irremediably. 
Note:

  the  detector  is  not  able  to  detect  gas  leaks  occurring  outside  the  room  where  it  is  installed,  neither  inside 

walls or under the floor. 

Warning

:  Consider  that  in  polluted  environments,  or  with  vapours  of  flammable  substances  (including  solvents), 

the lifetime of the sensor can be reduced.  Some substances cause a 

permanent reduction of sensitivity

; avoid 

contacts of the sensor with 

vapours of Silicon

 (found in paints, sealants and greases), Tetraethyl lead and Phos-

phate  esters.  Other  substances  cause  a  temporary  loss  of sensitivity;  these  "inhibitors"  include  Halogens,  Hydro-
gen sulphide, Chlorine, Chlorinated hydrocarbons (Trichloroethylene or Carbon tetrachloride).  After a short time in 
fresh air, the sensor resumes its normal operation.

 

CALIBRATION 

IMPORTANT NOTE: the following operations must be performed only by expert and trained personnel, the 
mA output changing value, causes the activation of alarm devices connected to the Control Unit. 

Periodical testing

: The Calibration with Gas/Air mixture is the only safe method to check the effective operation of 

the detector and the correct response of the detector to the gas. It is advisable to perform the  verification of opera-
tion every 6-12 months and every 2 years, if necessary, also proceed to recalibration with Gas/Air mixture. 

Operation Check, Zero Adjust, Calibration Check and Full Calibration

:

 are different code protected functions. 

To access these functions, it is necessary to open the housing (

only after setting the plant in safety

) and then  insert 

the  relevant  "

Code

"  with  the  F1  and  F2  keys.  To  have  the  key  pressure  recognized,  hold  pressing  it  for  around  a 

second (

until the Green Led doesn't switch off for a moment

). Then the next key can be pressed. In case of error, 

the Yellow LED flashes quickly, just wait about 10 seconds and the sequence is automatically cancelled. 

Calibration Kit and Sample Gas Bottles

 (

for Calibration Check and Full Calibration

) please, only using a mixture: 

20%LFL (0.88%v/v)

 

Methane in Air (20.9% Oxygen).

 

Catalytic sensors  cannot work in oxygen deficiency

. It  is possible  to use  either the  disposable cylinders with 

adjust valve or the high pressure ones with reduction gear.  It is also necessary to use 

the calibration kit 

Tecno-

control model TC011 

(

for non-corrosive gases

). 

“INSTRUMENT OPERATION CHECK” (

Check Code

: F2, F2, F1, F1)

this function allows to effect a functional test 

of the equipment. After having put the system in safety and inserted the "

Code Test

": 

STATUS of 

OPERATION CHECK

 

YELLOW LED GREEN LED  mA OUTPUT (S) 

Control Unit 

After entering the Check Code

 

OFF 

OFF 

0 mA 

FAULT 

Starting Operation Check

 

ON  

ON 

0 mA 

for 2 seconds

 

FAULT 

Within 2 sec. the mA output  

OFF 

ON 

increases to 4 mA for 5 seconds

 

NORM 

Within 2 sec. the output in mA 

ON 

ON 

increases to 

20 mA

 for 5 seconds

 

ALARM 

End Check and return to normal operation 

OFF 

ON 

gas measurement 

NORM 

The check lasts a total time of about 16 seconds, then the detector will return to normal operating conditions. 
It is advisable to perform this operation every 6-12 months according to the use. 

Note: 

The operation cannot be performed if the output is greater than 12mA (10% LFL). 

“ZERO ADJUST” (

Zero Code

: F2, F1, F1, F2)

even if there is an automatic zero adjustment system (zero tracker) 

this operation 

must be performed only if the output is different from 4mA

 (0% LFL on the control unit) after the 

installation and every 6-12 months based on the environmental conditions. This function must be done in clean air 
only (

environment without the presence of gas or other pollutants

). Immediately after having inserted the "

Zero Code

", as a con-

firmation of the operation carried out there will be 

1

 flash of the Green Led and the output will become 4,0mA. 

Note: 

The operation cannot be performed if the output is greater than 12mA (10%  LFL).

 In this case, it will be 

necessary to recalibrate the unit and/or replacing the detector.

 

“CALIBRATION CHECK” (

Cal Check Code

:  F2, F1, F2, F1)

:

 

this  operation  allow  to  effect  a  real  functional  test  of 

the equipment with gas

 

after the "

Calibration

" routine, or after the installation. This routine should be done during the 

periodic maintenances because this is the only method to verify the instrument real functioning. 

WARNING: the Cal Check Code allows using the same cylinder 0.88% v/v M Air for all types even 

if they have different gas response, 

e.g. the SE183KG. Instead, if a cylinder with 0.28% v/v  Air mixture 

is used for the verification, the code must not be used. 

1) 

With the keys perform the “

Cal Check Code

 ". Wait until the Yellow Led starts to flash (the green one remain fix) 

2) 

 Insert the TC011 over the sensor holder. Adjust the sample gas bottle valve (

0.88%v/v methane in air

), as the flow 

meter indicates around 0.3 l/mins (

see Fig.3

).  

3) 

Verify with a voltmeter connected to the Test-Point, the value reaches a value between 

184 and 216mV

, [cor-

responding to 

20mA

 (±1.6) output and the control unit should be display about 

20% LFL

 (±2)]. If the value re-

sult is more different, it is necessary to recalibrate the sensor (

see “

Calibration”

).  

Summary of Contents for SE183K

Page 1: ...00 max Sensore Sensor Capteur Catalitico Catalityc Catalytiques Campo di misura Standard range Champ de mesure 0 20 LFL LII ex LIE LEL Limite del Sensore Sensor limits Limite échelle 50 LFL LII Preriscaldo Preheating Préchauffage 60 secondi seconds secondes Tempo di risposta Response time Temps de résponse T90 60 secondi seconds secondes Precisione Accuracy Precision 10 2 LFL LII Deriva a lungo te...

Page 2: ...to si consiglia di verificare in aria pulita che l uscita sia circa 4 mA 0 LFL sulla centrale cui è collegato il rilevatore e solo se necessario procedere alla Regolazio ne del 4mA come indicato avanti in VERIFICA FUNZIONAMENTO Guasti in caso di guasto del sensore l uscita S va a 0 mA FAULT Il segnale sarà poi interpretato dalla centrale e segnalato come una situazione di guasto Nota i LED non son...

Page 3: ...perdite che avvengono fuori del locale in cui è installato o all interno dei muri o sotto il pavimento ATTENZIONE Considerare che in ambienti particolarmente inquinati o con vapori di sostanze infiammabili in par ticolare solventi potrebbe essere necessario eseguire più frequentemente la verifica e o la taratura periodica inol tre la vita utile del sensore può ridursi notevolmente Alcune sostanze ...

Page 4: ...enza di gas in fiammabili o altri inquinanti e usando solo la miscela Gas Metano al 20 LFL 0 88 v v in aria 20 9 O2 circa 1 Dopo aver eseguito il Codice Calibrazione il Led Verde inizierà a lampeggiare 2 Attendere che il Led Giallo si accenda fisso 3 Infilare tenerlo appoggiato il TC011 sul porta sensore e regolare l afflusso del Gas 0 88 v v Metano Aria in modo che il flussometro indichi circa 0 ...

Page 5: ...safe method to check the effective operation of the detector and the correct response to the gas It is recommended to perform the testing every 6 12 months Af ter 2 year we advise to perform the recalibration of the circuit with Gas Air mixture as explained later in the para graph FUNTIONALTESTING INSTALLATION The detector must be accurately installed according to all the national dispositions in ...

Page 6: ...cancelled Calibration Kit and Sample Gas Bottles for Calibration Check and Full Calibration please only using a mixture 20 LFL 0 88 v v Methane in Air 20 9 Oxygen Catalytic sensors cannot work in oxygen deficiency It is possible to use either the disposable cylinders with adjust valve or the high pressure ones with reduction gear It is also necessary to use the calibration kit Tecno control model ...

Page 7: ...ion catalytique étalonné jusque 20 de la LII Limite Inférieure d Inflammabilité La sonde est constituée par un boîtier antidéflagrant contenant le circuit électronique et les borniers de raccorde ment le capteur est inséré dans le porte capteur cylindrique placé dans la partie inférieure du boîtier Il s utilise en se raccordant sur les centrales d alarmes mono et multivoies Tecnocontrol indiqué au...

Page 8: ...n de gaz Évitez de monter le détecteur dans les coins près des portes ou des fenêtres ou dans le flux direct des prises d air ou des systèmes de ventilation Normalement le détecteur doit être installé à l intérieur mais s il est à l extérieur il doit être protégé de la lumière directe du soleil et de la pluie et éviter les sauts thermiques rapides Entrées de câble Le boîtier dans la partie supérie...

Page 9: ...FF ON Mesure de gaz NORM Le test électrique dure environ 16 secondes puis le détecteur revient à des conditions de fonctionnement nor males Il est souhaitable d exécuter cette opération tous les 6 12 mois en fonction de l utilisation Remarque Il n est pas possible d exécuter l opération si la sortie est supérieure à 12 mA 10 de LII RÉGLAGE DU ZÉRO Code du Zéro F2 F1 F1 F2 Bien que le détecteur dis...

Page 10: ... which can be connected to the SE183K series Centrales raccordable au SE183K Modello Model Modelé N Rilevatori Detectors number nombre de sondes SE127K SE128K SE139K ID250K 1 SE184K SE194K SE293K max 3 CE408 Max 8 CE424 max no 2 CE380UR Max 24 CE424 max 8 CE380UR no 1 max 8 CE380UR no 2 max 8 CE700 max no 23 CE380UR Max 184 Tabella 2 Table 2 Tableau 2 Sezione Cavo Cable Size Section du câble Resis...

Page 11: ...que EN 10226 1 Fissaggio fori di fissaggio Ø 6mm Fastening mounting holes Ø 6mm Fixation murale trous de fixation de Ø 6mm Ø42mm 100mm 75mm 65mm 128mm 165mm 0 3L min TC011 Fig 2 TC011 Tester di calibrazione Calibration Tester Tester d étalonnage Fig 1 Verifica Test Vérification Centralina Gas Gas Control Unit Centrale de détection S mV mV TEST 40 200mV F2 F1 TP2 S LED Verde Green Vert LED Giallo Y...

Reviews: