background image

IST-2126.KM03.01

 

Istruzione / User’s Manual / Manuel d’utilisation 

Pag.5/8

 

 

 

 

TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI ) 

 Tel: 02/26 92 28 90   Fax: 02/21 33 734

 

 

 

Model  SE126KM  positioning:

  Methane

  (CH

4

)  is  a  combustible  gas  lighter  than  air.  Its  density  related  to  air  is 

0.55; its LEL (

Lower Explosion Limit

) is 4.4% volume. 

The instrument should be fixed at 30 cm from the ceiling

 and be 

placed at 1-1.5 meter from the gas appliance, far from the room corners or from ventilation intakes. 

Model  SE126KG  positioning

:

  

LPG

 is a gas mixture composed by 20-30% Propane (C

3

H

8

) and by 70-80% Bu-

tane (C

4

H

10

). LPG calibration must be carried out with 

Butane

 which is a combustible gas heavier than air. Its den-

sity  related  to  air  is  2  and  its  LEL  (

Lower  Explosion  Limit

)  is  1.4%  volume. 

The instrument should be fixed at 30 cm. 

from the floor

 and be placed at 1-1.5 meter from the gas appliance, far from the room corners or from ventilation in-

takes 

Mounting

: The 

Fig. 2

 shows the instrument size. It has to be wall-mounted by four screw anchors. The 

normally 

closed manually resetting valve

 for the gas cut-off should be installed outside the room in a clearly indicated posi-

tion and should be protected from rain. 

Cable glands

: the lower side of the housing has 3 inputs designed for metric cable glands (

M20x1.5 that accept 

external cables Ø 6÷12 mm

). These passages are closed, but they are not 

manually breakable

, according to the in-

stallation requirements, they must be drilling. To facilitate the operation, they have a centering for the drill bit. 

Electrical Connection:

 

The installation must be provided for the mains, a bipolar disconnect switch dedicated for 

the gas detection system. The device, clearly identified, must act only on Phase and Neutral. If it is necessary, you 
may install a surge or lightning protector, etc. 

The instrument is supply with non-reversible and plug-in terminals. Therefore the cables should be anchored to the 
case in order to avoid terminal overstress. The connection to the detector does not need any hearting. 

Fig. 3

 shows the 230VAC powering connection with alarm siren and normally closed manual resetting valve. 

Fig. 4

 shows the 230VAC powering connection with alarm siren and normally open manual resetting valve. 

Fig. 5

 shows the 230VAC powering connection with alarm siren and our normally closed manual resetting valve 

with positioning sensor (models VR420÷VR480,) to verify, if an alarm occurs, the really gas cut-off. 

Fig. 5

 shows the 12VDC powering connection with both 12VDC alarm siren and normally closed manual resetting 

valve  with,  for  example,  our  power  supply  unit  PS175  (1,2A)  or  PS180  (2,5A)  with  internal  12VDC-7Ah  battery  to 
maintain the system powered on in absence of Main power supply. 

In this case, it is possible to connect the power 

supply also to 230VAC.

 

WARNING 

The detector 

doesn't need adjustments

 after being installed. 

Average life

the sensitive element used in this detector has an excellent stability in time. In fresh air and in normal 

working condition the sensor's life is more than 5 years from the date of installation. 

Important

: Catalytic sensors can operate only in presence of Oxygen. Do not use pure gases or a lighter directly 

on the sensor since they could damage it irremediably

. The detector is not able to detect gas leaks occurring out-

side the room where it is installed, neither inside walls nor under the floor. 

Attention

:  please  note  that  in  polluted  environments,  where  vapours  of  flammable  agents,  especially  solvents, 

might be present, the sensor's life can be reduced. Some substances cause a permanent reduction in sensitivity. 
Avoid  contacts  of  the  sensor  with  vapours  of  Silicone  compounds  (paintings  and  sealing paste), Tetra-ethyl Lead 
and  Phosphate  esters.  Some  other  substances  produce  a  temporary  loss  of sensitivity. These “inhibitors” include 
Hydrogen sulphide, Chlorine, Chlorinated hydrocarbons and halogenated compounds. The sensitivity is recovered 
after a short period of running in clear air. 

FUNTIONAL TESTING 

Periodical testing

we advise to carry out working tests every 6-12 months. 

PAY  ATTENTION

:  This  procedure  has  to  be  made  with extreme attention and by authorized and trained people; 

because starting this procedure it will start both Outputs (relays) causing the activation of connected alarm devices. 
The 

TEST

 key is deactivating when an alarm condition occurs. 

Instrument operation check

: push the 

TEST

 button for 3 seconds until all Led and relays deactivates.  

Then  the  LED  will  illuminate  in  sequence.  With  the  LEDs  the  corresponding  relays  will  activate  (Yellow  LED  and 
FAULT  relay,  2

nd

  red  LED and PREAL. Relay, 3

rd

 red LED and ALARM relay). After 5 seconds the instrument re-

turns to normal working condition. NOTE: during the Test some conditions occur: 

All 5 LEDs illuminates

ONLY

 when the detector is connected with a control wire for certain types of valves. In this 

case the test results correct. 

Yellow  LED off (after red LEDs)

: indicates the lack of control wire of the valve or that the valve is closed. If this 

wire is disconnected from the detector the test results correct, if not please check the valve and repeat test. If test 
result is unchanged, electrical problems are in place, please contact the supplier.  

The  3rd  Red  LED  off

ONLY 

when the detector is connected with a control wire for certain types of valves. The 

control wire is disconnected or the valve is open. Please check both the valve status and the wire connection and 
repeat the test. If wire is disconnected, check if the wire on the terminal “AUX” is connected and repeat the test. 

All other cases: 

If some LEDs do not light on or some relays do not switch, the detector is damaged. In this case 

please replace it or send it back to the supplier. 

Calibration testing with gas mixture

the calibration is carried out with specific gas in our laboratory. We advise 

to check the calibration only 

using sample gas bottles with a mixture about 20% or 25%LEL Gas in Air (

20.9% 

Oxygen

) (Gas Methane for SE126KM and Butane for SE126KG). Catalytic sensors cannot work without Oxy-

gen

. The calibration testing should be done with Calibration Kit model TC011 (See specific user’s manual). Connect 

the  sample  gas  bottle  to  the  flow  meter,  open  Gas  (0.3  l/min  flow  rate),  wait  for  3  minutes  and  check  that  the  in-

Summary of Contents for SE126KM

Page 1: ...il d intervention de alarme 1 10 LIE LEL Intervento Allarme 2 2nd Alarm intervention Seuil d intervention de alarme 2 20 LIE LEL Contatti relè Contacts rating Contact relais 230Vac 3A SPDT Tempo di risposta T90 Response Time T90 Temps de réponse T90 30 secondi seconds secondes Temperatura Umidità di funzionamento Operation Temp Humidity Temp et humidité de fonctionnement 10 50 C 5 90 RH non conden...

Page 2: ...re di posizione e nel caso in cui essa sia scattata si spegnerà il Led giallo e si disattiverà il relè FAULT vedi sezione Guasti più avanti Ripristino dopo un Allarme la centralina tornerà al funzionamento normale solo se la concentrazione di gas è in feriore al 10 del LIE il relè ALARM sarà disattivato il 3 e 4 Led rosso si spegneranno Se attivato sarà disattiva to il relè FAULT e si spegnerà il ...

Page 3: ...ori del locale in cui è installato o all interno dei muri o sotto il pavimento ATTENZIONE Considerare che in ambienti particolarmente inquinati o con vapori di sostanze infiammabili in par ticolare solventi la vita utile del sensore può ridursi notevolmente Alcune sostanze causano una riduzione perma nente di sensibilità Evitare che il sensore venga a contatto con vapori di Silicone presente in ve...

Page 4: ...sor mod VR420 VR480 connected to AUX input in the case that the gas is still open the yellow LED and the FAULT Relay will activate see FAULTS section RESET Key The alarm condition remains latched LEDs and relays remain activated even if the gas concentra tion is reduced because the mounted valve is closed To reset the normal working conditions and or switch off the siren push the RESET key Buzzer ...

Page 5: ...than 5 years from the date of installation Important Catalytic sensors can operate only in presence of Oxygen Do not use pure gases or a lighter directly on the sensor since they could damage it irremediably The detector is not able to detect gas leaks occurring out side the room where it is installed neither inside walls nor under the floor Attention please note that in polluted environments wher...

Page 6: ... ma nuelle sur le bouton RESET Cette action ne sera possible que si l appareil à ce moment ne détecte pas de pré sence gazeuse Arrêt de la Sirène d alarme si la concentration de gaz est supérieure 10 de la LIE le relais PREAL serait dé sactivée et le deuxième LED rouge clignote Apres 20 seconds les deux revient in préalarme la sirène sonne Si la vanne NF à réarmement manuel avec senseur de fin de ...

Page 7: ...tion Vie moyenne ce capteur a une excellente stabilité dans le temps En condition de fonctionnement normale en air non pollué la vie utile du capteur est supérieure de 5 ans à partir de la date de mise sous tension Attention Les capteurs à combustion catalytique fonctionnent seulement en présence d Oxygène Eviter l usage de gaz pur ou du briquet sur le capteur qui pourrait être irrémédiablement en...

Page 8: ...A NC C NA NC C NA L N SE301A 12Vdc PREAL ALARM Valvola NC 12Vcc 12VDC NC Valve Vanne n f 12Vcc FAULT AUX 230Vca PS180 2 5A 7Ah 12Vdc 230Vca Fig 5 Collegamento con valvola a riarmo manuale Norm Chiusa Wiring diagram with a normally closed manual resetting valve Schéma avec vanne à réarmement manuel norm fermée 230Vca NC C NA NC C NA NC C NA L N SE301A PREAL ALARM VR420 VR480 FAULT AUX Fig 4 Collega...

Reviews: