TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date: 09/05/06
83
Final Control Neumovent Graph Ventilator
Operation
Description
110
PCV (Pressure Control) Mode
Select PCV
Programming:
PCV = 10 ± 2 cm H2O
Ti= 1.5 sec.
Peep = 0
The rest is programmed with <CTRL> + <ENTER>
Check: Peak Pressure Reading = 10 ± 2 cm H2O
Expired VT reading = 500 ml ± 8%
120
PSV Mode
Select PSV
a) Program Backup (Backup Ventilation).
Apnea = 15 sec.
SIMV = ON
Confirm other backup parameters with <CTRL> + <ENTER>
b) Program PSV.
Pressure = 15 cm. H2O
Peep = 0
ENTER
Trigger the test lung.
Check the mode operation (with the graph generated on the screen).
Check Peak Pressure = 15 cm +/- 2 cm H2O
Apnea Alarm
Do not trigger the test lung.
At 15 sec., the apnea alarm should become activated.
The ventilator turns to backup ventilation.
Trigger the test lung.
Check if it returns to PSV ventilation.
130
SIMV VCV Mode
Programming: f= 6 per minute. Inspiratory time = 1 sec.
PSV 15 cm H2O
The other parameters remain the same as for the initial programming.
ENTER
Trigger the test lung.
Check the operation of the ventilatory mode.
140
SIMV PC Mode
Programming: f= 6 per minute. Inspiratory time = 1 sec.
PCV = 20 cm H2O
PSV = 15 cm H2O
<ENTER>.
Trigger the test lung.
Check the operation of the ventilatory mode.
Summary of Contents for NEUMOVENT
Page 1: ...Graph ventilator Technical manual review 04 MAY 2006 ...
Page 3: ...CHAPTER I GENERAL AND OPERATION FEATURES ...
Page 12: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 9 ...
Page 13: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 10 ...
Page 30: ...CHAPTER II MAINTENANCE INSTRUCTIONS ...
Page 32: ...CHAPTER III TROUBLE SHOOTING ...
Page 37: ...CHAPTER IV SENSOR VERIFICATION ...
Page 44: ...CHAPTER V EQUIPMENT OPENING AND CLOSURE ...
Page 60: ...CHAPTER VI DETAIL OF ASSEMBLIES ...
Page 70: ...CHAPTER VII ELECTRONIC BOARDS DETAIL ...
Page 78: ...CHAPTER VIII CALIBRATION ...
Page 90: ...CHAPTER IX FINAL CONTROL ...