TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date: 09/05/06
34
Sensor Inspection
-
The value for the opening of the chosen valve may be increased or decreased with selection keys
This opening value may range between 0 (Closed Valve) and a maximum of 4095 (Valve completely
opened). If press
“
ctrl
”
+
“
selection
”
the values go up or dawn in tens.
-
All valves (Opening value = 0) may be closed with the
“
Reset
”
key.
The
“AirVal”
line corresponds to the air proportional valve.
The
“
Oxy
Val”
line corresponds to the oxygen proportional valve.
The
“SuffVal”
line corresponds to the antisofocation valve
The
“
PeepVal
”
line corresponds to the PEEP proportional valve.
Air and Oxygen Valves
(AirVal and OxyValv)
When the opening value is increased with a value ranging between 0 and 4095, we will notice that the
valve remains closed while the opening value is lower than 900 units (approximately). When the
opening exceeds this value, the valve begins to be opened progressively (air begins to flow through the
inspiratory outlet) until the maximum value is reached (greater than 160 L/min), i.e., when the opening
value reaches 4095 units.
The valve can then be closed using the
“
Reset
”
.
Sensors Check
In the upper left corner of the window, there is an 8-line column.
Each lines shows the digitalized readings of the flow and pressure. These readings will vary depending
on the flows or pressures being applied on the sensors.
When there are no supply gases connected to the device, the readings given by these sensors are
the ones shown in the following table:
The second column
shows a value equal to the first one
but multiplied by four (do not take this into account).
Air Flow:
0091
0364
Oxy Flow:
0080
0320
Right press:
0390
1562
Left Press:
0441
1767
Exp Flow:
0069
0278
Oxy Source:
0930
0900
Air Source:
0930
Power Source: 0881
3525
Oxy cell:
0000
0000
Table :1
Summary of Contents for NEUMOVENT
Page 1: ...Graph ventilator Technical manual review 04 MAY 2006 ...
Page 3: ...CHAPTER I GENERAL AND OPERATION FEATURES ...
Page 12: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 9 ...
Page 13: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 10 ...
Page 30: ...CHAPTER II MAINTENANCE INSTRUCTIONS ...
Page 32: ...CHAPTER III TROUBLE SHOOTING ...
Page 37: ...CHAPTER IV SENSOR VERIFICATION ...
Page 44: ...CHAPTER V EQUIPMENT OPENING AND CLOSURE ...
Page 60: ...CHAPTER VI DETAIL OF ASSEMBLIES ...
Page 70: ...CHAPTER VII ELECTRONIC BOARDS DETAIL ...
Page 78: ...CHAPTER VIII CALIBRATION ...
Page 90: ...CHAPTER IX FINAL CONTROL ...