17

H B - 1 D   O P E R A T O R  ’ S   M A N U A L

The HB-1D

1

Insert the tubes or tube holder into the HB-1D ensuring that they rotate freely and
engage on the drive rollers.

2

Close the door.

3

Connect the mains electricity supply to the mains socket at the rear of the unit.
Switch the power ON from the switch located next to the mains input lead.  The
POWER indicator and the display on the front panel will light.

4

The Heater Indicator will be lit while the heater is on.

Summary of Contents for Barloworld Scientific HB-1D

Page 1: ...HB 1D Hybridiser Operator s Manual Issue 25 Date of issue 3rd May 2005 Techne is a trademark Techne 2003...

Page 2: ......

Page 3: ...itial checks 9 Specification 10 Unpacking 11 Installation 11 Warning 13 Servicing by Techne and their Agents 13 Switching on 14 Temperature control 14 Speed control 14 OPERATION The tubes 15 The HB 1D...

Page 4: ...2 H B 1 D O P E R A T O R S M A N U A L...

Page 5: ...3 H B 1 D O P E R A T O R S M A N U A L NOTES...

Page 6: ...1785 813748 e mail equipment barloworld scientific com or Techne Inc 3 Terri Lane Suite 10 Burlington New Jersey 08016 USA Telephone 609 589 2560 Toll free 800 225 9243 ext 306 Fax 609 589 2571 e mai...

Page 7: ...le remplacement ne peut tre effectu que par un personnel qualifi Si les fusibles sautent sans arr t il s agit d un probl me s rieux Nous vous conseillons dans ce cas de prendre contact avec votre four...

Page 8: ...brennt liegt ein gr erer Defekt vor Das Ger t mu zur Reparatur an Ihren Lieferanten eingesandt werden Einleitung Bitte lesen Sie diese Bedienungsanleitung komplett bevor Sie dieses Ger t benutzen Warn...

Page 9: ...de muestras recuerde que las piezas del equipo tales como tubos bloques y dem s accesorios pueden estar muy calientes Tome las precauciones mencionadas anteriormente Garant a Este aparato est garantiz...

Page 10: ...ne Inc 3 Terri Lane Suite 10 Burlington New Jersey 08016 USA Telephone 609 589 2560 Toll free 800 225 9243 ext 306 Fax 609 589 2571 e mail labproducts techneusa com Web site www techneusa com Introduz...

Page 11: ...it will hold 4 8 12 or 16 mini tubes if you use a four tube carrier e the HB 1D comes with adjustable feet to enable accurate levelling Initial Checks When you receive your new HB 1D remove all the pa...

Page 12: ...W Temperature selection Digital Temperature Display LED Temperature Sensor PRT External Dimensions Width 385mm Depth 285mm Height 555mm Internal Dimensions Width 260mm Depth 210mm Height 330mm Weight...

Page 13: ...ra Tubes if ordered especially Installation 1 Place the unit on a suitable bench a flat workspace or in a safety cupboard if required ensuring that the air inlet and outlet louvres are not obstructed...

Page 14: ...12 H B 1 D O P E R A T O R S M A N U A L SPEED CONTROL...

Page 15: ...when the spindle cap is removed DO NOT TIP THE TUBE SO THAT THE COLOURED END IS DOWN AS THIS WILL CAUSE A SPILLAGE 3 The unit has been designed to conform to international safety specifications and i...

Page 16: ...button Temperature Set button Controller Display When you first switch on the display will show the edition of the software which your unit uses For example software issue 1 would be shown as follows...

Page 17: ...the end cap A small quantity of Techne O ring grease see the list of accessories must be smeared on the seal to facilitate the insertion of the end cap Remove the spindle cap and the liquids appropri...

Page 18: ...t different ends of the carrier Fix the tube into a multi tube holder using the springs and ensure that there are three tubes in the holder The multi tube holder will not rotate properly if there are...

Page 19: ...y and engage on the drive rollers 2 Close the door 3 Connect the mains electricity supply to the mains socket at the rear of the unit Switch the power ON from the switch located next to the mains inpu...

Page 20: ...perature when pressed The UP ARROW Button When the SET temperature button is held down and the UP ARROW button is pressed the set temperature is increased The DOWN ARROW Button When the SET temperatur...

Page 21: ...ow the set temperature If the light is on continuously the heater is getting constant power The only exception is described under Over Temperature Indicator As the temperature approaches the set tempe...

Page 22: ...inverting the tube Add hybridisation liquid with radioactive probe added by using a syringe and needle through the hole in the blue end cap The minimum volume for a single filter is 5ml WASHES Washing...

Page 23: ...entre of the button Repeated operation of the cutout indicates a serious fault you may need to return your unit to your supplier for repair Fuses If the power indicator fails to light when power is su...

Page 24: ...it from the power supply Follow this sequence of operations when the HB 1D has been allowed to cool below at least 50 C a Remove all glassware from the HB 1D b Remove the drip tray c Clean all surface...

Page 25: ...Grease tube FHB32 44mm external diameter hybridisation tube assembly complete with seal and screw on end cap FHB31 44mm external diameter tube only 6007625 Seal for use with 44mm external diameter tub...

Page 26: ...24 H B 1 D O P E R A T O R S M A N U A L...

Page 27: ...l 41 41 541 5877 Fax 41 41 541 5878 Email General Information Sales info kesn ch Accounts Department accounts kesn ch USA O ce 233 Rogue River Highway 425 Grants Pass Oregon 97527 5429 USA Tel 1 503 2...