![Techkon SpectroDens Manual Download Page 50](http://html1.mh-extra.com/html/techkon/spectrodens/spectrodens_manual_1066520050.webp)
50
Manual TECHKON SpectroDens
The measurement conditions M0, M1, M2, M3 according to
ISO 13655-2009
Many types of paper contain optical brighteners or pigments to improve the visual
performance of the paper. The brighteners and pigments are used to make the paper
look brighter and whiter. But for the accurate color adjustment they are quite a chal-
lenge, because the type of ambient light influences the color perception in most cases
to a large extent. To be able to evaluate these printing substrates precisely when mea-
suring, new standards have been defined concerning the measurement conditions.
The following measurement conditions are selectable in the measurement parameter
menu of the SpectroDens color measurement device.
Measurement condition M0
is used when the printing substrate contains no or only
little optical brighteners. The measurement light corresponds to the standard illumina-
tion type A and the UV component of the source, which activates the optical brighten-
ers, is not specified. Almost every color measurement device in the graphic industries
works so far with this measurement condition. This measurement condition should
also be used, when measurement values have to be compliant to older devices.
Measurement condition M1
was defined to activate optical brighteners with a speci-
fied component of UV light, to be able to evaluate the effect of the brighteners more
precisely. The measurement light conforms to the illumination type D 50 with a speci-
fied UV component.
Measurement condition M2
describes the measurement without UV light. In older
devices this measurement condition was realized using an UV-Cut filter. The compari-
son of the measurement value M1 to M2 shows clearly how strong the effect of optical
brighteners is.
Measurement condition M3
is mainly used for densitometric measurements and only
to a less extent for colorimetric measurements. It describes the use of polarization
filters which reduce surface reflections to minimize measurement deviations between
wet and dry printing sheets.